Evaluation of the New Digital Goldmann Applanation Tonometer for Measuring Intraocular Pressure
المؤلفون المشاركون
Sakaue, Yuta
Ueda, Jun
Seki, Masaaki
Tanaka, Takayuki
Togano, Tetsuya
Yoshino, Takaiko
Fukuchi, Takeo
المصدر
العدد
المجلد 2014، العدد 2014 (31 ديسمبر/كانون الأول 2014)، ص ص. 1-5، 5ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2014-07-10
دولة النشر
مصر
عدد الصفحات
5
التخصصات الرئيسية
الملخص EN
Purpose.
To compare a new digital Goldmann applanation tonometer (dGAT) that measures intraocular pressure (IOP) in 0.1 mmHg increments to a standard Goldmann applanation tonometer (sGAT).
Methods.
This study included 116 eyes of 60 subjects.
A single examiner first measured IOP in triplicate using either sGAT or dGAT, which was randomly chosen.
After a 5-minute interval, the next set of three consecutive IOP was measured using the other GAT.
Results.
The mean IOP measured with sGAT was 16.27 ± 6.68 mmHg and 16.35 ± 6.69 mmHg with dGAT.
Pearson’s correlation coefficient was 0.998 ( P < 0.01 ).
The subjects were divided into three groups based on the mean IOP: IOP < 14 mmHg, 14–20 mmHg, or >20 mmHg.
The Pearson’s correlation coefficient within each group was 0.935, 0.972, and 0.997 ( P < 0.01 ), respectively.
The difference within the three consecutive IOP measurements (maximum–minimum) for dGAT (0.72 ± 0.34 mmHg) was significantly smaller than those with sGAT (0.92 ± 0.42 mmHg, P < 0.01 ).
Even in patients with equal IOP (zero left-right difference) with sGAT ( n = 30 ), dGAT detected IOP differences between the left and right eyes (0.47 ± 0.31 mmHg).
Conclusion.
Compared to sGAT, dGAT measurements are highly reproducible and less variable.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Sakaue, Yuta& Ueda, Jun& Seki, Masaaki& Tanaka, Takayuki& Togano, Tetsuya& Yoshino, Takaiko…[et al.]. 2014. Evaluation of the New Digital Goldmann Applanation Tonometer for Measuring Intraocular Pressure. Journal of Ophthalmology،Vol. 2014, no. 2014, pp.1-5.
https://search.emarefa.net/detail/BIM-1042524
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Sakaue, Yuta…[et al.]. Evaluation of the New Digital Goldmann Applanation Tonometer for Measuring Intraocular Pressure. Journal of Ophthalmology No. 2014 (2014), pp.1-5.
https://search.emarefa.net/detail/BIM-1042524
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Sakaue, Yuta& Ueda, Jun& Seki, Masaaki& Tanaka, Takayuki& Togano, Tetsuya& Yoshino, Takaiko…[et al.]. Evaluation of the New Digital Goldmann Applanation Tonometer for Measuring Intraocular Pressure. Journal of Ophthalmology. 2014. Vol. 2014, no. 2014, pp.1-5.
https://search.emarefa.net/detail/BIM-1042524
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-1042524
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
![](/images/ebook-kashef.png)
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر
![](/images/kashef-image.png)