Evaluation of the New Digital Goldmann Applanation Tonometer for Measuring Intraocular Pressure

Joint Authors

Sakaue, Yuta
Ueda, Jun
Seki, Masaaki
Tanaka, Takayuki
Togano, Tetsuya
Yoshino, Takaiko
Fukuchi, Takeo

Source

Journal of Ophthalmology

Issue

Vol. 2014, Issue 2014 (31 Dec. 2014), pp.1-5, 5 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2014-07-10

Country of Publication

Egypt

No. of Pages

5

Main Subjects

Medicine

Abstract EN

Purpose.

To compare a new digital Goldmann applanation tonometer (dGAT) that measures intraocular pressure (IOP) in 0.1 mmHg increments to a standard Goldmann applanation tonometer (sGAT).

Methods.

This study included 116 eyes of 60 subjects.

A single examiner first measured IOP in triplicate using either sGAT or dGAT, which was randomly chosen.

After a 5-minute interval, the next set of three consecutive IOP was measured using the other GAT.

Results.

The mean IOP measured with sGAT was 16.27 ± 6.68 mmHg and 16.35 ± 6.69 mmHg with dGAT.

Pearson’s correlation coefficient was 0.998 ( P < 0.01 ).

The subjects were divided into three groups based on the mean IOP: IOP < 14 mmHg, 14–20 mmHg, or >20 mmHg.

The Pearson’s correlation coefficient within each group was 0.935, 0.972, and 0.997 ( P < 0.01 ), respectively.

The difference within the three consecutive IOP measurements (maximum–minimum) for dGAT (0.72 ± 0.34 mmHg) was significantly smaller than those with sGAT (0.92 ± 0.42 mmHg, P < 0.01 ).

Even in patients with equal IOP (zero left-right difference) with sGAT ( n = 30 ), dGAT detected IOP differences between the left and right eyes (0.47 ± 0.31 mmHg).

Conclusion.

Compared to sGAT, dGAT measurements are highly reproducible and less variable.

American Psychological Association (APA)

Sakaue, Yuta& Ueda, Jun& Seki, Masaaki& Tanaka, Takayuki& Togano, Tetsuya& Yoshino, Takaiko…[et al.]. 2014. Evaluation of the New Digital Goldmann Applanation Tonometer for Measuring Intraocular Pressure. Journal of Ophthalmology،Vol. 2014, no. 2014, pp.1-5.
https://search.emarefa.net/detail/BIM-1042524

Modern Language Association (MLA)

Sakaue, Yuta…[et al.]. Evaluation of the New Digital Goldmann Applanation Tonometer for Measuring Intraocular Pressure. Journal of Ophthalmology No. 2014 (2014), pp.1-5.
https://search.emarefa.net/detail/BIM-1042524

American Medical Association (AMA)

Sakaue, Yuta& Ueda, Jun& Seki, Masaaki& Tanaka, Takayuki& Togano, Tetsuya& Yoshino, Takaiko…[et al.]. Evaluation of the New Digital Goldmann Applanation Tonometer for Measuring Intraocular Pressure. Journal of Ophthalmology. 2014. Vol. 2014, no. 2014, pp.1-5.
https://search.emarefa.net/detail/BIM-1042524

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-1042524