Liquid Chromatography-Tandem Mass Spectrometric Assay for Determination of Stavudine in Human Plasma
المؤلف
المصدر
العدد
المجلد 2014، العدد 2014 (31 ديسمبر/كانون الأول 2014)، ص ص. 1-5، 5ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2014-06-17
دولة النشر
مصر
عدد الصفحات
5
التخصصات الرئيسية
الملخص EN
A LC-MS/MS method for determination of stavudine in human plasma was established and validated, and it was applied to the pharmaceutical formulations bioequivalence study.
0.5 mL plasma sample was extracted by liquid-liquid extraction.
Stavudine was detected by a LC-MS/MS system.
The pharmacokinetic parameters of stavudine in different formulations were calculated by noncompartment model statistics.
The method was linear over the concentration ranges 5.00–1000 ng/mL in plasma.
The intra- and interassay relative standard deviation (RSD) was <10%.
The average accuracies for the assay at three concentrations (5.00, 80.0, and 900 ng/mL) were from 100.2% to 102.5%.
Pharmacokinetic parameters of stavudine reference formulation were obtained as follows: Tmax was 0.6±0.2 h, Cmax was 480.7±150.9 g/L, t1/2 was 1.7±0.4 h, and AUC0-t was 872.8±227.8 g·h/L, and pharmacokinetic parameters of stavudine test formulation were obtained as follows: Tmax was 0.5±0.2 h, Cmax was 537.5±178.5 g/L, t1/2 was 1.7±0.3 h, and AUC0-t was (914.1±284.5) g·h/L.
Calculated with AUC0-t, the bioavailability of two formulations was 105.0%.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Jin, Fengdan. 2014. Liquid Chromatography-Tandem Mass Spectrometric Assay for Determination of Stavudine in Human Plasma. Journal of Spectroscopy،Vol. 2014, no. 2014, pp.1-5.
https://search.emarefa.net/detail/BIM-1043103
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Jin, Fengdan. Liquid Chromatography-Tandem Mass Spectrometric Assay for Determination of Stavudine in Human Plasma. Journal of Spectroscopy No. 2014 (2014), pp.1-5.
https://search.emarefa.net/detail/BIM-1043103
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Jin, Fengdan. Liquid Chromatography-Tandem Mass Spectrometric Assay for Determination of Stavudine in Human Plasma. Journal of Spectroscopy. 2014. Vol. 2014, no. 2014, pp.1-5.
https://search.emarefa.net/detail/BIM-1043103
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-1043103
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر