Comprasion of ICare Rebound Tonometer and Goldmann Applanation Tonometer in High Myopia

المؤلفون المشاركون

Cagatay, Halil Huseyin
Yazar, Zeliha
Ekinci, Metin
Gokce, Gokcen
Ceylan, Erdinc

المصدر

The Scientific World Journal

العدد

المجلد 2014، العدد 2014 (31 ديسمبر/كانون الأول 2014)، ص ص. 1-5، 5ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2014-10-14

دولة النشر

مصر

عدد الصفحات

5

التخصصات الرئيسية

الطب البشري
تكنولوجيا المعلومات وعلم الحاسوب

الملخص EN

Purpose.

To compare intraocular pressure (IOP) measurements with the Goldmann applanation tonometer (GAT) and the ICare rebound tonometer (RBT) in high myopic eyes.

Patients and Methods.

This randomized prospective study included 40 eyes of 40 patients with high myopia.

All patients’ central corneal thickness (CCT), anterior chamber depth (ACD), axial length (AXL), keratometry, and refractive measurements were recorded and followed by IOP measurement with RBT and GAT.

Results.

The average CCT, AXL, and ACD were determined to be 514.65±32 μm, 27.65±2.22 mm, and 3.25±0.51 mm, respectively.

Mean K was 43.27±1.4 D and mean spherical equivalent was -11.31±4.30 D.

The mean IOP values obtained by RBT and GAT were 17.18±3.72 mmHg and 16.48±3.19 mmHg, respectively.

The deviations of RBT readings from corrected GAT values were highly correlated with CCT values (r=0.588, P=0.0001).

The mean corrected GAT reading was 17.49±3.01 mmHg.

Linear regression analysis showed that a CCT change of 10 μm resulted in an RBT reading deviation of 0.57 mmHg.

The Bland-Altman scatter-plot and McNemar test showed a clinically good level of agreement between the two tonometers.

Conclusion.

This study found a good agreement level between the two tonometers in high myopic patients and that RBT measurements are influenced by CCT variations.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Cagatay, Halil Huseyin& Ekinci, Metin& Yazar, Zeliha& Gokce, Gokcen& Ceylan, Erdinc. 2014. Comprasion of ICare Rebound Tonometer and Goldmann Applanation Tonometer in High Myopia. The Scientific World Journal،Vol. 2014, no. 2014, pp.1-5.
https://search.emarefa.net/detail/BIM-1051419

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Cagatay, Halil Huseyin…[et al.]. Comprasion of ICare Rebound Tonometer and Goldmann Applanation Tonometer in High Myopia. The Scientific World Journal No. 2014 (2014), pp.1-5.
https://search.emarefa.net/detail/BIM-1051419

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Cagatay, Halil Huseyin& Ekinci, Metin& Yazar, Zeliha& Gokce, Gokcen& Ceylan, Erdinc. Comprasion of ICare Rebound Tonometer and Goldmann Applanation Tonometer in High Myopia. The Scientific World Journal. 2014. Vol. 2014, no. 2014, pp.1-5.
https://search.emarefa.net/detail/BIM-1051419

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1051419