Comprasion of ICare Rebound Tonometer and Goldmann Applanation Tonometer in High Myopia

Joint Authors

Cagatay, Halil Huseyin
Yazar, Zeliha
Ekinci, Metin
Gokce, Gokcen
Ceylan, Erdinc

Source

The Scientific World Journal

Issue

Vol. 2014, Issue 2014 (31 Dec. 2014), pp.1-5, 5 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2014-10-14

Country of Publication

Egypt

No. of Pages

5

Main Subjects

Medicine
Information Technology and Computer Science

Abstract EN

Purpose.

To compare intraocular pressure (IOP) measurements with the Goldmann applanation tonometer (GAT) and the ICare rebound tonometer (RBT) in high myopic eyes.

Patients and Methods.

This randomized prospective study included 40 eyes of 40 patients with high myopia.

All patients’ central corneal thickness (CCT), anterior chamber depth (ACD), axial length (AXL), keratometry, and refractive measurements were recorded and followed by IOP measurement with RBT and GAT.

Results.

The average CCT, AXL, and ACD were determined to be 514.65±32 μm, 27.65±2.22 mm, and 3.25±0.51 mm, respectively.

Mean K was 43.27±1.4 D and mean spherical equivalent was -11.31±4.30 D.

The mean IOP values obtained by RBT and GAT were 17.18±3.72 mmHg and 16.48±3.19 mmHg, respectively.

The deviations of RBT readings from corrected GAT values were highly correlated with CCT values (r=0.588, P=0.0001).

The mean corrected GAT reading was 17.49±3.01 mmHg.

Linear regression analysis showed that a CCT change of 10 μm resulted in an RBT reading deviation of 0.57 mmHg.

The Bland-Altman scatter-plot and McNemar test showed a clinically good level of agreement between the two tonometers.

Conclusion.

This study found a good agreement level between the two tonometers in high myopic patients and that RBT measurements are influenced by CCT variations.

American Psychological Association (APA)

Cagatay, Halil Huseyin& Ekinci, Metin& Yazar, Zeliha& Gokce, Gokcen& Ceylan, Erdinc. 2014. Comprasion of ICare Rebound Tonometer and Goldmann Applanation Tonometer in High Myopia. The Scientific World Journal،Vol. 2014, no. 2014, pp.1-5.
https://search.emarefa.net/detail/BIM-1051419

Modern Language Association (MLA)

Cagatay, Halil Huseyin…[et al.]. Comprasion of ICare Rebound Tonometer and Goldmann Applanation Tonometer in High Myopia. The Scientific World Journal No. 2014 (2014), pp.1-5.
https://search.emarefa.net/detail/BIM-1051419

American Medical Association (AMA)

Cagatay, Halil Huseyin& Ekinci, Metin& Yazar, Zeliha& Gokce, Gokcen& Ceylan, Erdinc. Comprasion of ICare Rebound Tonometer and Goldmann Applanation Tonometer in High Myopia. The Scientific World Journal. 2014. Vol. 2014, no. 2014, pp.1-5.
https://search.emarefa.net/detail/BIM-1051419

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-1051419