![](/images/graphics-bg.png)
Does Metal Transfer Differ on Retrieved Ceramic and CoCr Femoral Heads?
المؤلفون المشاركون
Lee, Gwo-Chin
Kurtz, Steven M.
Fredette, Eliza K.
MacDonald, Daniel W.
Underwood, Richard J.
Chen, Antonia F.
Mont, Michael A.
Klein, Gregg R.
Rimnac, Clare M.
المصدر
العدد
المجلد 2015، العدد 2015 (31 ديسمبر/كانون الأول 2015)، ص ص. 1-10، 10ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2015-10-25
دولة النشر
مصر
عدد الصفحات
10
التخصصات الرئيسية
الملخص EN
Metal transfer has been observed on retrieved THA femoral heads for both CoCr and ceramic bearing materials.
In vitro wear testing has shown increased wear to polyethylene acetabular liners with the presence of metal transfer.
This study sought to investigate the extent of metal transfer on the bearing surface of CoCr and ceramic femoral heads and identify prevalent morphologies.
Three bearing couple cohorts: M-PE ( n = 50 ), C-PE ( n = 35 ), and C-C ( n = 15 ), were derived from two previously matched collections ( n = 50 /group) of CoCr and ceramic femoral heads.
From the three cohorts, 75% of the femoral heads showed visual evidence of metal transfer.
These femoral heads were analyzed using direct measurement, digital photogrammetry, and white light interferometry.
Surface area coverage and curved median surface area were similar among the three cohorts.
The most prevalent metal transfer patterns observed were random stripes ( n = 21 / 75 ), longitudinal stripes ( n = 17 / 75 ), and random patches ( n = 13 / 75 ).
Metal transfer arc length was shorter in the M-PE cohort.
Understanding the morphology of metal transfer may be useful for more realistic recreation of metal transfer in in vitro pin-on-disk and joint simulators studies.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Fredette, Eliza K.& MacDonald, Daniel W.& Underwood, Richard J.& Chen, Antonia F.& Mont, Michael A.& Lee, Gwo-Chin…[et al.]. 2015. Does Metal Transfer Differ on Retrieved Ceramic and CoCr Femoral Heads?. BioMed Research International،Vol. 2015, no. 2015, pp.1-10.
https://search.emarefa.net/detail/BIM-1054876
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Fredette, Eliza K.…[et al.]. Does Metal Transfer Differ on Retrieved Ceramic and CoCr Femoral Heads?. BioMed Research International No. 2015 (2015), pp.1-10.
https://search.emarefa.net/detail/BIM-1054876
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Fredette, Eliza K.& MacDonald, Daniel W.& Underwood, Richard J.& Chen, Antonia F.& Mont, Michael A.& Lee, Gwo-Chin…[et al.]. Does Metal Transfer Differ on Retrieved Ceramic and CoCr Femoral Heads?. BioMed Research International. 2015. Vol. 2015, no. 2015, pp.1-10.
https://search.emarefa.net/detail/BIM-1054876
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-1054876
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
![](/images/ebook-kashef.png)
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر
![](/images/kashef-image.png)