Does Metal Transfer Differ on Retrieved Ceramic and CoCr Femoral Heads?

Joint Authors

Lee, Gwo-Chin
Kurtz, Steven M.
Fredette, Eliza K.
MacDonald, Daniel W.
Underwood, Richard J.
Chen, Antonia F.
Mont, Michael A.
Klein, Gregg R.
Rimnac, Clare M.

Source

BioMed Research International

Issue

Vol. 2015, Issue 2015 (31 Dec. 2015), pp.1-10, 10 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2015-10-25

Country of Publication

Egypt

No. of Pages

10

Main Subjects

Medicine

Abstract EN

Metal transfer has been observed on retrieved THA femoral heads for both CoCr and ceramic bearing materials.

In vitro wear testing has shown increased wear to polyethylene acetabular liners with the presence of metal transfer.

This study sought to investigate the extent of metal transfer on the bearing surface of CoCr and ceramic femoral heads and identify prevalent morphologies.

Three bearing couple cohorts: M-PE ( n = 50 ), C-PE ( n = 35 ), and C-C ( n = 15 ), were derived from two previously matched collections ( n = 50 /group) of CoCr and ceramic femoral heads.

From the three cohorts, 75% of the femoral heads showed visual evidence of metal transfer.

These femoral heads were analyzed using direct measurement, digital photogrammetry, and white light interferometry.

Surface area coverage and curved median surface area were similar among the three cohorts.

The most prevalent metal transfer patterns observed were random stripes ( n = 21 / 75 ), longitudinal stripes ( n = 17 / 75 ), and random patches ( n = 13 / 75 ).

Metal transfer arc length was shorter in the M-PE cohort.

Understanding the morphology of metal transfer may be useful for more realistic recreation of metal transfer in in vitro pin-on-disk and joint simulators studies.

American Psychological Association (APA)

Fredette, Eliza K.& MacDonald, Daniel W.& Underwood, Richard J.& Chen, Antonia F.& Mont, Michael A.& Lee, Gwo-Chin…[et al.]. 2015. Does Metal Transfer Differ on Retrieved Ceramic and CoCr Femoral Heads?. BioMed Research International،Vol. 2015, no. 2015, pp.1-10.
https://search.emarefa.net/detail/BIM-1054876

Modern Language Association (MLA)

Fredette, Eliza K.…[et al.]. Does Metal Transfer Differ on Retrieved Ceramic and CoCr Femoral Heads?. BioMed Research International No. 2015 (2015), pp.1-10.
https://search.emarefa.net/detail/BIM-1054876

American Medical Association (AMA)

Fredette, Eliza K.& MacDonald, Daniel W.& Underwood, Richard J.& Chen, Antonia F.& Mont, Michael A.& Lee, Gwo-Chin…[et al.]. Does Metal Transfer Differ on Retrieved Ceramic and CoCr Femoral Heads?. BioMed Research International. 2015. Vol. 2015, no. 2015, pp.1-10.
https://search.emarefa.net/detail/BIM-1054876

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-1054876