![](/images/graphics-bg.png)
Does Metal Transfer Differ on Retrieved Ceramic and CoCr Femoral Heads?
Joint Authors
Lee, Gwo-Chin
Kurtz, Steven M.
Fredette, Eliza K.
MacDonald, Daniel W.
Underwood, Richard J.
Chen, Antonia F.
Mont, Michael A.
Klein, Gregg R.
Rimnac, Clare M.
Source
Issue
Vol. 2015, Issue 2015 (31 Dec. 2015), pp.1-10, 10 p.
Publisher
Hindawi Publishing Corporation
Publication Date
2015-10-25
Country of Publication
Egypt
No. of Pages
10
Main Subjects
Abstract EN
Metal transfer has been observed on retrieved THA femoral heads for both CoCr and ceramic bearing materials.
In vitro wear testing has shown increased wear to polyethylene acetabular liners with the presence of metal transfer.
This study sought to investigate the extent of metal transfer on the bearing surface of CoCr and ceramic femoral heads and identify prevalent morphologies.
Three bearing couple cohorts: M-PE ( n = 50 ), C-PE ( n = 35 ), and C-C ( n = 15 ), were derived from two previously matched collections ( n = 50 /group) of CoCr and ceramic femoral heads.
From the three cohorts, 75% of the femoral heads showed visual evidence of metal transfer.
These femoral heads were analyzed using direct measurement, digital photogrammetry, and white light interferometry.
Surface area coverage and curved median surface area were similar among the three cohorts.
The most prevalent metal transfer patterns observed were random stripes ( n = 21 / 75 ), longitudinal stripes ( n = 17 / 75 ), and random patches ( n = 13 / 75 ).
Metal transfer arc length was shorter in the M-PE cohort.
Understanding the morphology of metal transfer may be useful for more realistic recreation of metal transfer in in vitro pin-on-disk and joint simulators studies.
American Psychological Association (APA)
Fredette, Eliza K.& MacDonald, Daniel W.& Underwood, Richard J.& Chen, Antonia F.& Mont, Michael A.& Lee, Gwo-Chin…[et al.]. 2015. Does Metal Transfer Differ on Retrieved Ceramic and CoCr Femoral Heads?. BioMed Research International،Vol. 2015, no. 2015, pp.1-10.
https://search.emarefa.net/detail/BIM-1054876
Modern Language Association (MLA)
Fredette, Eliza K.…[et al.]. Does Metal Transfer Differ on Retrieved Ceramic and CoCr Femoral Heads?. BioMed Research International No. 2015 (2015), pp.1-10.
https://search.emarefa.net/detail/BIM-1054876
American Medical Association (AMA)
Fredette, Eliza K.& MacDonald, Daniel W.& Underwood, Richard J.& Chen, Antonia F.& Mont, Michael A.& Lee, Gwo-Chin…[et al.]. Does Metal Transfer Differ on Retrieved Ceramic and CoCr Femoral Heads?. BioMed Research International. 2015. Vol. 2015, no. 2015, pp.1-10.
https://search.emarefa.net/detail/BIM-1054876
Data Type
Journal Articles
Language
English
Notes
Includes bibliographical references
Record ID