Validation and Development of a New Automatic Algorithm for Time-Resolved Segmentation of the Left Ventricle in Magnetic Resonance Imaging
المؤلفون المشاركون
Tufvesson, Jane
Hedström, Erik
Steding-Ehrenborg, Katarina
Carlsson, Marcus
Arheden, Håkan
Heiberg, Einar
المصدر
العدد
المجلد 2015، العدد 2015 (31 ديسمبر/كانون الأول 2015)، ص ص. 1-12، 12ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2015-06-21
دولة النشر
مصر
عدد الصفحات
12
التخصصات الرئيسية
الملخص EN
Introduction.
Manual delineation of the left ventricle is clinical standard for quantification of cardiovascular magnetic resonance images despite being time consuming and observer dependent.
Previous automatic methods generally do not account for one major contributor to stroke volume, the long-axis motion.
Therefore, the aim of this study was to develop and validate an automatic algorithm for time-resolved segmentation covering the whole left ventricle, including basal slices affected by long-axis motion.
Methods.
Ninety subjects imaged with a cine balanced steady state free precession sequence were included in the study (training set n = 40 , test set n = 50 ).
Manual delineation was reference standard and second observer analysis was performed in a subset ( n = 25 ).
The automatic algorithm uses deformable model with expectation-maximization, followed by automatic removal of papillary muscles and detection of the outflow tract.
Results.
The mean differences between automatic segmentation and manual delineation were EDV −11 mL, ESV 1 mL, EF −3%, and LVM 4 g in the test set.
Conclusions.
The automatic LV segmentation algorithm reached accuracy comparable to interobserver for manual delineation, thereby bringing automatic segmentation one step closer to clinical routine.
The algorithm and all images with manual delineations are available for benchmarking.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Tufvesson, Jane& Hedström, Erik& Steding-Ehrenborg, Katarina& Carlsson, Marcus& Arheden, Håkan& Heiberg, Einar. 2015. Validation and Development of a New Automatic Algorithm for Time-Resolved Segmentation of the Left Ventricle in Magnetic Resonance Imaging. BioMed Research International،Vol. 2015, no. 2015, pp.1-12.
https://search.emarefa.net/detail/BIM-1057433
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Tufvesson, Jane…[et al.]. Validation and Development of a New Automatic Algorithm for Time-Resolved Segmentation of the Left Ventricle in Magnetic Resonance Imaging. BioMed Research International No. 2015 (2015), pp.1-12.
https://search.emarefa.net/detail/BIM-1057433
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Tufvesson, Jane& Hedström, Erik& Steding-Ehrenborg, Katarina& Carlsson, Marcus& Arheden, Håkan& Heiberg, Einar. Validation and Development of a New Automatic Algorithm for Time-Resolved Segmentation of the Left Ventricle in Magnetic Resonance Imaging. BioMed Research International. 2015. Vol. 2015, no. 2015, pp.1-12.
https://search.emarefa.net/detail/BIM-1057433
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-1057433
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
![](/images/ebook-kashef.png)
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر
![](/images/kashef-image.png)