An Adaptive Thresholding Method for BTV Estimation Incorporating PET Reconstruction Parameters: A Multicenter Study of the Robustness and the Reliability

المؤلفون المشاركون

Botta, F.
Fioroni, F.
Grassi, E.
Pacilio, M.
Cremonesi, M.
Brambilla, M.
Matheoud, R.
Basile, C.
Bracco, C.
Cavedon, C.
Morzenti, S.
Giri, M.
Gallivanone, F.
De Ponti, E.
Stasi, M.
Pasetto, S.
Valzano, S.
Zanni, D.
Castiglioni, Isabella

المصدر

Computational and Mathematical Methods in Medicine

العدد

المجلد 2015، العدد 2015 (31 ديسمبر/كانون الأول 2015)، ص ص. 1-12، 12ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2015-05-19

دولة النشر

مصر

عدد الصفحات

12

التخصصات الرئيسية

الطب البشري

الملخص EN

Objective.

The aim of this work was to assess robustness and reliability of an adaptive thresholding algorithm for the biological target volume estimation incorporating reconstruction parameters.

Method.

In a multicenter study, a phantom with spheres of different diameters (6.5–57.4 mm) was filled with 18F-FDG at different target-to-background ratios (TBR: 2.5–70) and scanned for different acquisition periods (2–5 min).

Image reconstruction algorithms were used varying number of iterations and postreconstruction transaxial smoothing.

Optimal thresholds (TS) for volume estimation were determined as percentage of the maximum intensity in the cross section area of the spheres.

Multiple regression techniques were used to identify relevant predictors of TS.

Results.

The goodness of the model fit was high (R2: 0.74–0.92).

TBR was the most significant predictor of TS.

For all scanners, except the Gemini scanners, FWHM was an independent predictor of TS.

Significant differences were observed between scanners of different models, but not between different scanners of the same model.

The shrinkage on cross validation was small and indicative of excellent reliability of model estimation.

Conclusions.

Incorporation of postreconstruction filtering FWHM in an adaptive thresholding algorithm for the BTV estimation allows obtaining a robust and reliable method to be applied to a variety of different scanners, without scanner-specific individual calibration.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Brambilla, M.& Matheoud, R.& Basile, C.& Bracco, C.& Castiglioni, Isabella& Cavedon, C.…[et al.]. 2015. An Adaptive Thresholding Method for BTV Estimation Incorporating PET Reconstruction Parameters: A Multicenter Study of the Robustness and the Reliability. Computational and Mathematical Methods in Medicine،Vol. 2015, no. 2015, pp.1-12.
https://search.emarefa.net/detail/BIM-1057938

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Brambilla, M.…[et al.]. An Adaptive Thresholding Method for BTV Estimation Incorporating PET Reconstruction Parameters: A Multicenter Study of the Robustness and the Reliability. Computational and Mathematical Methods in Medicine No. 2015 (2015), pp.1-12.
https://search.emarefa.net/detail/BIM-1057938

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Brambilla, M.& Matheoud, R.& Basile, C.& Bracco, C.& Castiglioni, Isabella& Cavedon, C.…[et al.]. An Adaptive Thresholding Method for BTV Estimation Incorporating PET Reconstruction Parameters: A Multicenter Study of the Robustness and the Reliability. Computational and Mathematical Methods in Medicine. 2015. Vol. 2015, no. 2015, pp.1-12.
https://search.emarefa.net/detail/BIM-1057938

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1057938