Revision of a Fractured Uncemented Revision Stem Using a Custom Designed Punch and Retrograde Through-Knee Approach

المؤلفون المشاركون

Nasr, P. J.
Keene, G. S.

المصدر

Case Reports in Orthopedics

العدد

المجلد 2015، العدد 2015 (31 ديسمبر/كانون الأول 2015)، ص ص. 1-3، 3ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2015-02-22

دولة النشر

مصر

عدد الصفحات

3

التخصصات الرئيسية

الطب البشري

الملخص EN

We report a unique case of a fractured modular cobalt chromium connection taper Revitan (Zimmer, Warsaw, IN) revision prosthesis.

Macroscopic examination revealed a fracture at the diaphyseal-metaphyseal junction of this modular component.

This report highlights that fractures can still occur with modern modular prostheses.

We are not aware of any published failures of the Revitan revision prosthesis.

We also describe a unique method of retrieval for a broken well fixed uncemented femoral stem, using a custom designed extraction instrument via a through-knee approach.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Nasr, P. J.& Keene, G. S.. 2015. Revision of a Fractured Uncemented Revision Stem Using a Custom Designed Punch and Retrograde Through-Knee Approach. Case Reports in Orthopedics،Vol. 2015, no. 2015, pp.1-3.
https://search.emarefa.net/detail/BIM-1059490

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Nasr, P. J.& Keene, G. S.. Revision of a Fractured Uncemented Revision Stem Using a Custom Designed Punch and Retrograde Through-Knee Approach. Case Reports in Orthopedics No. 2015 (2015), pp.1-3.
https://search.emarefa.net/detail/BIM-1059490

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Nasr, P. J.& Keene, G. S.. Revision of a Fractured Uncemented Revision Stem Using a Custom Designed Punch and Retrograde Through-Knee Approach. Case Reports in Orthopedics. 2015. Vol. 2015, no. 2015, pp.1-3.
https://search.emarefa.net/detail/BIM-1059490

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1059490