Revision of a Fractured Uncemented Revision Stem Using a Custom Designed Punch and Retrograde Through-Knee Approach

Joint Authors

Nasr, P. J.
Keene, G. S.

Source

Case Reports in Orthopedics

Issue

Vol. 2015, Issue 2015 (31 Dec. 2015), pp.1-3, 3 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2015-02-22

Country of Publication

Egypt

No. of Pages

3

Main Subjects

Medicine

Abstract EN

We report a unique case of a fractured modular cobalt chromium connection taper Revitan (Zimmer, Warsaw, IN) revision prosthesis.

Macroscopic examination revealed a fracture at the diaphyseal-metaphyseal junction of this modular component.

This report highlights that fractures can still occur with modern modular prostheses.

We are not aware of any published failures of the Revitan revision prosthesis.

We also describe a unique method of retrieval for a broken well fixed uncemented femoral stem, using a custom designed extraction instrument via a through-knee approach.

American Psychological Association (APA)

Nasr, P. J.& Keene, G. S.. 2015. Revision of a Fractured Uncemented Revision Stem Using a Custom Designed Punch and Retrograde Through-Knee Approach. Case Reports in Orthopedics،Vol. 2015, no. 2015, pp.1-3.
https://search.emarefa.net/detail/BIM-1059490

Modern Language Association (MLA)

Nasr, P. J.& Keene, G. S.. Revision of a Fractured Uncemented Revision Stem Using a Custom Designed Punch and Retrograde Through-Knee Approach. Case Reports in Orthopedics No. 2015 (2015), pp.1-3.
https://search.emarefa.net/detail/BIM-1059490

American Medical Association (AMA)

Nasr, P. J.& Keene, G. S.. Revision of a Fractured Uncemented Revision Stem Using a Custom Designed Punch and Retrograde Through-Knee Approach. Case Reports in Orthopedics. 2015. Vol. 2015, no. 2015, pp.1-3.
https://search.emarefa.net/detail/BIM-1059490

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-1059490