Accuracy and Precision of Three-Dimensional Low Dose CT Compared to Standard RSA in Acetabular Cups: An Experimental Study

المؤلفون المشاركون

Maguire, Gerald Q.
Zeleznik, Michael P.
Brodén, Cyrus
Olivecrona, Henrik
Noz, Marilyn E.
Sköldenberg, Olof

المصدر

BioMed Research International

العدد

المجلد 2016، العدد 2016 (31 ديسمبر/كانون الأول 2016)، ص ص. 1-6، 6ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2016-07-10

دولة النشر

مصر

عدد الصفحات

6

التخصصات الرئيسية

الطب البشري

الملخص EN

Background and Purpose.

The gold standard for detection of implant wear and migration is currently radiostereometry (RSA).

The purpose of this study is to compare a three-dimensional computed tomography technique (3D CT) to standard RSA as an alternative technique for measuring migration of acetabular cups in total hip arthroplasty.

Materials and Methods.

With tantalum beads, we marked one cemented and one uncemented cup and mounted these on a similarly marked pelvic model.

A comparison was made between 3D CT and standard RSA for measuring migration.

Twelve repeated stereoradiographs and CT scans with double examinations in each position and gradual migration of the implants were made.

Precision and accuracy of the 3D CT were calculated.

Results.

The accuracy of the 3D CT ranged between 0.07 and 0.32 mm for translations and 0.21 and 0.82° for rotation.

The precision ranged between 0.01 and 0.09 mm for translations and 0.06 and 0.29° for rotations, respectively.

For standard RSA, the precision ranged between 0.04 and 0.09 mm for translations and 0.08 and 0.32° for rotations, respectively.

There was no significant difference in precision between 3D CT and standard RSA.

The effective radiation dose of the 3D CT method, comparable to RSA, was estimated to be 0.33 mSv.

Interpretation.

Low dose 3D CT is a comparable method to standard RSA in an experimental setting.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Brodén, Cyrus& Olivecrona, Henrik& Maguire, Gerald Q.& Noz, Marilyn E.& Zeleznik, Michael P.& Sköldenberg, Olof. 2016. Accuracy and Precision of Three-Dimensional Low Dose CT Compared to Standard RSA in Acetabular Cups: An Experimental Study. BioMed Research International،Vol. 2016, no. 2016, pp.1-6.
https://search.emarefa.net/detail/BIM-1098255

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Brodén, Cyrus…[et al.]. Accuracy and Precision of Three-Dimensional Low Dose CT Compared to Standard RSA in Acetabular Cups: An Experimental Study. BioMed Research International No. 2016 (2016), pp.1-6.
https://search.emarefa.net/detail/BIM-1098255

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Brodén, Cyrus& Olivecrona, Henrik& Maguire, Gerald Q.& Noz, Marilyn E.& Zeleznik, Michael P.& Sköldenberg, Olof. Accuracy and Precision of Three-Dimensional Low Dose CT Compared to Standard RSA in Acetabular Cups: An Experimental Study. BioMed Research International. 2016. Vol. 2016, no. 2016, pp.1-6.
https://search.emarefa.net/detail/BIM-1098255

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1098255