Accuracy and Precision of Three-Dimensional Low Dose CT Compared to Standard RSA in Acetabular Cups: An Experimental Study

Joint Authors

Maguire, Gerald Q.
Zeleznik, Michael P.
Brodén, Cyrus
Olivecrona, Henrik
Noz, Marilyn E.
Sköldenberg, Olof

Source

BioMed Research International

Issue

Vol. 2016, Issue 2016 (31 Dec. 2016), pp.1-6, 6 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2016-07-10

Country of Publication

Egypt

No. of Pages

6

Main Subjects

Medicine

Abstract EN

Background and Purpose.

The gold standard for detection of implant wear and migration is currently radiostereometry (RSA).

The purpose of this study is to compare a three-dimensional computed tomography technique (3D CT) to standard RSA as an alternative technique for measuring migration of acetabular cups in total hip arthroplasty.

Materials and Methods.

With tantalum beads, we marked one cemented and one uncemented cup and mounted these on a similarly marked pelvic model.

A comparison was made between 3D CT and standard RSA for measuring migration.

Twelve repeated stereoradiographs and CT scans with double examinations in each position and gradual migration of the implants were made.

Precision and accuracy of the 3D CT were calculated.

Results.

The accuracy of the 3D CT ranged between 0.07 and 0.32 mm for translations and 0.21 and 0.82° for rotation.

The precision ranged between 0.01 and 0.09 mm for translations and 0.06 and 0.29° for rotations, respectively.

For standard RSA, the precision ranged between 0.04 and 0.09 mm for translations and 0.08 and 0.32° for rotations, respectively.

There was no significant difference in precision between 3D CT and standard RSA.

The effective radiation dose of the 3D CT method, comparable to RSA, was estimated to be 0.33 mSv.

Interpretation.

Low dose 3D CT is a comparable method to standard RSA in an experimental setting.

American Psychological Association (APA)

Brodén, Cyrus& Olivecrona, Henrik& Maguire, Gerald Q.& Noz, Marilyn E.& Zeleznik, Michael P.& Sköldenberg, Olof. 2016. Accuracy and Precision of Three-Dimensional Low Dose CT Compared to Standard RSA in Acetabular Cups: An Experimental Study. BioMed Research International،Vol. 2016, no. 2016, pp.1-6.
https://search.emarefa.net/detail/BIM-1098255

Modern Language Association (MLA)

Brodén, Cyrus…[et al.]. Accuracy and Precision of Three-Dimensional Low Dose CT Compared to Standard RSA in Acetabular Cups: An Experimental Study. BioMed Research International No. 2016 (2016), pp.1-6.
https://search.emarefa.net/detail/BIM-1098255

American Medical Association (AMA)

Brodén, Cyrus& Olivecrona, Henrik& Maguire, Gerald Q.& Noz, Marilyn E.& Zeleznik, Michael P.& Sköldenberg, Olof. Accuracy and Precision of Three-Dimensional Low Dose CT Compared to Standard RSA in Acetabular Cups: An Experimental Study. BioMed Research International. 2016. Vol. 2016, no. 2016, pp.1-6.
https://search.emarefa.net/detail/BIM-1098255

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID