A Case of Implant Failure in Partial Wrist Fusion Applying Magnesium-Based Headless Bone Screws

المؤلفون المشاركون

Mittlmeier, Thomas
Wichelhaus, Alice
Emmerich, Judith

المصدر

Case Reports in Orthopedics

العدد

المجلد 2016، العدد 2016 (31 ديسمبر/كانون الأول 2016)، ص ص. 1-5، 5ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2016-10-05

دولة النشر

مصر

عدد الصفحات

5

التخصصات الرئيسية

الطب البشري

الملخص EN

This article presents a case of implant failure resulting in mechanical instability of a scaphotrapezotrapezoideal arthrodesis using magnesium-based headless bone screws.

During revision surgery osteolysis surrounding the screws was observed as well as degraded screw threads already in existence at 6 weeks after implantation.

The supposed osseous integration attributed to magnesium-based screws could not be reproduced in this particular case.

Thus, it can be reasoned that the use of magnesium-based screws for partial wrist arthrodesis cannot be encouraged, at least not in dual use.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Wichelhaus, Alice& Emmerich, Judith& Mittlmeier, Thomas. 2016. A Case of Implant Failure in Partial Wrist Fusion Applying Magnesium-Based Headless Bone Screws. Case Reports in Orthopedics،Vol. 2016, no. 2016, pp.1-5.
https://search.emarefa.net/detail/BIM-1102050

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Wichelhaus, Alice…[et al.]. A Case of Implant Failure in Partial Wrist Fusion Applying Magnesium-Based Headless Bone Screws. Case Reports in Orthopedics No. 2016 (2016), pp.1-5.
https://search.emarefa.net/detail/BIM-1102050

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Wichelhaus, Alice& Emmerich, Judith& Mittlmeier, Thomas. A Case of Implant Failure in Partial Wrist Fusion Applying Magnesium-Based Headless Bone Screws. Case Reports in Orthopedics. 2016. Vol. 2016, no. 2016, pp.1-5.
https://search.emarefa.net/detail/BIM-1102050

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1102050