![](/images/graphics-bg.png)
A Case of Implant Failure in Partial Wrist Fusion Applying Magnesium-Based Headless Bone Screws
Joint Authors
Mittlmeier, Thomas
Wichelhaus, Alice
Emmerich, Judith
Source
Issue
Vol. 2016, Issue 2016 (31 Dec. 2016), pp.1-5, 5 p.
Publisher
Hindawi Publishing Corporation
Publication Date
2016-10-05
Country of Publication
Egypt
No. of Pages
5
Main Subjects
Abstract EN
This article presents a case of implant failure resulting in mechanical instability of a scaphotrapezotrapezoideal arthrodesis using magnesium-based headless bone screws.
During revision surgery osteolysis surrounding the screws was observed as well as degraded screw threads already in existence at 6 weeks after implantation.
The supposed osseous integration attributed to magnesium-based screws could not be reproduced in this particular case.
Thus, it can be reasoned that the use of magnesium-based screws for partial wrist arthrodesis cannot be encouraged, at least not in dual use.
American Psychological Association (APA)
Wichelhaus, Alice& Emmerich, Judith& Mittlmeier, Thomas. 2016. A Case of Implant Failure in Partial Wrist Fusion Applying Magnesium-Based Headless Bone Screws. Case Reports in Orthopedics،Vol. 2016, no. 2016, pp.1-5.
https://search.emarefa.net/detail/BIM-1102050
Modern Language Association (MLA)
Wichelhaus, Alice…[et al.]. A Case of Implant Failure in Partial Wrist Fusion Applying Magnesium-Based Headless Bone Screws. Case Reports in Orthopedics No. 2016 (2016), pp.1-5.
https://search.emarefa.net/detail/BIM-1102050
American Medical Association (AMA)
Wichelhaus, Alice& Emmerich, Judith& Mittlmeier, Thomas. A Case of Implant Failure in Partial Wrist Fusion Applying Magnesium-Based Headless Bone Screws. Case Reports in Orthopedics. 2016. Vol. 2016, no. 2016, pp.1-5.
https://search.emarefa.net/detail/BIM-1102050
Data Type
Journal Articles
Language
English
Notes
Includes bibliographical references
Record ID