A Case of Implant Failure in Partial Wrist Fusion Applying Magnesium-Based Headless Bone Screws

Joint Authors

Mittlmeier, Thomas
Wichelhaus, Alice
Emmerich, Judith

Source

Case Reports in Orthopedics

Issue

Vol. 2016, Issue 2016 (31 Dec. 2016), pp.1-5, 5 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2016-10-05

Country of Publication

Egypt

No. of Pages

5

Main Subjects

Medicine

Abstract EN

This article presents a case of implant failure resulting in mechanical instability of a scaphotrapezotrapezoideal arthrodesis using magnesium-based headless bone screws.

During revision surgery osteolysis surrounding the screws was observed as well as degraded screw threads already in existence at 6 weeks after implantation.

The supposed osseous integration attributed to magnesium-based screws could not be reproduced in this particular case.

Thus, it can be reasoned that the use of magnesium-based screws for partial wrist arthrodesis cannot be encouraged, at least not in dual use.

American Psychological Association (APA)

Wichelhaus, Alice& Emmerich, Judith& Mittlmeier, Thomas. 2016. A Case of Implant Failure in Partial Wrist Fusion Applying Magnesium-Based Headless Bone Screws. Case Reports in Orthopedics،Vol. 2016, no. 2016, pp.1-5.
https://search.emarefa.net/detail/BIM-1102050

Modern Language Association (MLA)

Wichelhaus, Alice…[et al.]. A Case of Implant Failure in Partial Wrist Fusion Applying Magnesium-Based Headless Bone Screws. Case Reports in Orthopedics No. 2016 (2016), pp.1-5.
https://search.emarefa.net/detail/BIM-1102050

American Medical Association (AMA)

Wichelhaus, Alice& Emmerich, Judith& Mittlmeier, Thomas. A Case of Implant Failure in Partial Wrist Fusion Applying Magnesium-Based Headless Bone Screws. Case Reports in Orthopedics. 2016. Vol. 2016, no. 2016, pp.1-5.
https://search.emarefa.net/detail/BIM-1102050

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-1102050