The Effects of Sample Transport by Pneumatic Tube System on Routine Hematology and Coagulation Tests
المؤلفون المشاركون
Subbarayan, Devi
Choccalingam, Chidambharam
Lakshmi, Chittode Kodumudi Anantha
المصدر
العدد
المجلد 2018، العدد 2018 (31 ديسمبر/كانون الأول 2018)، ص ص. 1-4، 4ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2018-07-02
دولة النشر
مصر
عدد الصفحات
4
التخصصات الرئيسية
الملخص EN
Background.
Automation helps improve laboratory operational efficiency and reduce the turnaround time.
Pneumatic tube systems (PTS) automate specimen transport between the lab and other areas of the hospital.
Its effect on complete blood count (CBC) and coagulation is still controversial.
Aim.
To study the effects of pneumatic tube system sample transport on complete blood count and coagulation parameters to compare them with hand delivered samples.
Methods.
75 paired samples for complete blood count and 25 paired samples for coagulation analysis were compared between samples sent via pneumatic tube system and hand delivered system.
Results.
PTS showed significant decrease in red cell indices such as MCV and RDW and increase in MCHC.
Other red cell parameters and WBC parameters showed no statistical significant difference.
Statistically significant increase in platelet count was observed with PTS samples.
However, these differences were clinically insignificant.
No significant effect of PTS was found in PT and APTT samples compared to the hand delivered samples.
Conclusion.
Despite statistically significant changes in RBC parameters such as MCV, RDW, and MCHC and platelet count, these changes were clinically insignificant.
Hence, blood samples for CBC and coagulation assay can safely be transported via our hospital’s PTS.
However, further studies on platelet count are warranted to ensure safe transport and accuracy of the results.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Subbarayan, Devi& Choccalingam, Chidambharam& Lakshmi, Chittode Kodumudi Anantha. 2018. The Effects of Sample Transport by Pneumatic Tube System on Routine Hematology and Coagulation Tests. Advances in Hematology،Vol. 2018, no. 2018, pp.1-4.
https://search.emarefa.net/detail/BIM-1117753
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Subbarayan, Devi…[et al.]. The Effects of Sample Transport by Pneumatic Tube System on Routine Hematology and Coagulation Tests. Advances in Hematology No. 2018 (2018), pp.1-4.
https://search.emarefa.net/detail/BIM-1117753
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Subbarayan, Devi& Choccalingam, Chidambharam& Lakshmi, Chittode Kodumudi Anantha. The Effects of Sample Transport by Pneumatic Tube System on Routine Hematology and Coagulation Tests. Advances in Hematology. 2018. Vol. 2018, no. 2018, pp.1-4.
https://search.emarefa.net/detail/BIM-1117753
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-1117753
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
![](/images/ebook-kashef.png)
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر
![](/images/kashef-image.png)