Contrast Microsphere Destruction by a Continuous Flow Ventricular Assist Device: An In Vitro Evaluation Using a Mock Circulation Loop
المؤلفون المشاركون
Fraser, John F.
Platts, David G.
Bartnikowski, Nicole
Scalia, Gregory M.
Gregory, Shaun D.
المصدر
العدد
المجلد 2017، العدد 2017 (31 ديسمبر/كانون الأول 2017)، ص ص. 1-9، 9ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2017-08-13
دولة النشر
مصر
عدد الصفحات
9
التخصصات الرئيسية
الملخص EN
Objectives.
Transthoracic echocardiography (TTE) is fundamental in managing patients supported with ventricular assist devices (VAD).
However imaging can be difficult in these patients.
Contrast improves image quality but they are hydrodynamically fragile agents.
The aim was to assess contrast concentration following passage through a VAD utilising a mock circulation loop (MCL).
Methods.
Heartware continuous flow (CF) VAD was incorporated into a MCL.
Definity® contrast was infused into the MCL with imaging before and after CF-VAD.
5 mm2 regions of interest were used to obtain signal intensity (decibels), as a surrogate of contrast concentration.
Results.
Four pump speeds revealed significant reduction in contrast signal intensity after CF-VAD compared to before CF-VAD (all p<0.0001).
Combined pre- and postpump data at all speeds showed a 22.2% absolute reduction in contrast signal intensity across the CF-VAD (14.8 ± 0.8 dB prepump versus 11.6 ± 1.4 dB postpump; p<0.0001).
Mean signal intensity reduction at each speed showed an inverse relationship between speed and relative reduction in signal intensity.
Conclusion.
Contrast microsphere transit through a CF-VAD within a MCL resulted in significant reduction in signal intensity, consistent with destruction within the pump.
This was evident at all CF-VAD pump speeds but relative signal drop was inversely proportional to pump speed.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Platts, David G.& Bartnikowski, Nicole& Gregory, Shaun D.& Scalia, Gregory M.& Fraser, John F.. 2017. Contrast Microsphere Destruction by a Continuous Flow Ventricular Assist Device: An In Vitro Evaluation Using a Mock Circulation Loop. BioMed Research International،Vol. 2017, no. 2017, pp.1-9.
https://search.emarefa.net/detail/BIM-1137333
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Platts, David G.…[et al.]. Contrast Microsphere Destruction by a Continuous Flow Ventricular Assist Device: An In Vitro Evaluation Using a Mock Circulation Loop. BioMed Research International No. 2017 (2017), pp.1-9.
https://search.emarefa.net/detail/BIM-1137333
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Platts, David G.& Bartnikowski, Nicole& Gregory, Shaun D.& Scalia, Gregory M.& Fraser, John F.. Contrast Microsphere Destruction by a Continuous Flow Ventricular Assist Device: An In Vitro Evaluation Using a Mock Circulation Loop. BioMed Research International. 2017. Vol. 2017, no. 2017, pp.1-9.
https://search.emarefa.net/detail/BIM-1137333
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-1137333
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
![](/images/ebook-kashef.png)
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر
![](/images/kashef-image.png)