Development of an Analytical Protocol for Determination of Cyanide in Human Biological Samples Based on Application of Ion Chromatography with Pulsed Amperometric Detection

المؤلفون المشاركون

Ruman, Marek
Namieśnik, Jacek
Jaszczak, Ewa
Narkowicz, Sylwia
Polkowska, Żaneta

المصدر

Journal of Analytical Methods in Chemistry

العدد

المجلد 2017، العدد 2017 (31 ديسمبر/كانون الأول 2017)، ص ص. 1-7، 7ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2017-11-21

دولة النشر

مصر

عدد الصفحات

7

التخصصات الرئيسية

الكيمياء

الملخص EN

A simple and accurate ion chromatography (IC) method with pulsed amperometric detection (PAD) was proposed for the determination of cyanide ion in urine, sweat, and saliva samples.

The sample pretreatment relies on alkaline digestion and application of Dionex OnGuard II H cartridge.

Under the optimized conditions, the method showed good linearity in the range of 1–100 μg/L for urine, 5–100 μg/L for saliva, and 3–100 μg/L for sweat samples with determination coefficients (R)>0.992.

Low detection limits (LODs) in the range of 1.8 μg/L, 5.1 μg/L, and 5.8 μg/L for urine, saliva, and sweat samples, respectively, and good repeatability (CV < 3%, n=3) were obtained.

The proposed method has been successfully applied to the analysis of human biological samples.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Jaszczak, Ewa& Ruman, Marek& Narkowicz, Sylwia& Namieśnik, Jacek& Polkowska, Żaneta. 2017. Development of an Analytical Protocol for Determination of Cyanide in Human Biological Samples Based on Application of Ion Chromatography with Pulsed Amperometric Detection. Journal of Analytical Methods in Chemistry،Vol. 2017, no. 2017, pp.1-7.
https://search.emarefa.net/detail/BIM-1170287

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Jaszczak, Ewa…[et al.]. Development of an Analytical Protocol for Determination of Cyanide in Human Biological Samples Based on Application of Ion Chromatography with Pulsed Amperometric Detection. Journal of Analytical Methods in Chemistry No. 2017 (2017), pp.1-7.
https://search.emarefa.net/detail/BIM-1170287

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Jaszczak, Ewa& Ruman, Marek& Narkowicz, Sylwia& Namieśnik, Jacek& Polkowska, Żaneta. Development of an Analytical Protocol for Determination of Cyanide in Human Biological Samples Based on Application of Ion Chromatography with Pulsed Amperometric Detection. Journal of Analytical Methods in Chemistry. 2017. Vol. 2017, no. 2017, pp.1-7.
https://search.emarefa.net/detail/BIM-1170287

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل