An Evaluation of the Accuracy of a Flash Glucose Monitoring System in Children with Diabetes in comparison with Venous Blood Glucose
المؤلفون المشاركون
Cao, Bing-Yan
Gong, Chun-xiu
Wu, Di
Liu, Min
Li, Wenjing
Su, Chang
Liang, Xuejun
Wang, Rui
Chen, Jiajia
Yi, Yajun
المصدر
العدد
المجلد 2019، العدد 2019 (31 ديسمبر/كانون الأول 2019)، ص ص. 1-7، 7ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2019-09-09
دولة النشر
مصر
عدد الصفحات
7
التخصصات الرئيسية
الملخص EN
Aims.
To evaluate the performance of a factory-calibrated flash glucose monitoring system in children with diabetes compared to venous blood glucose (BG).
Methods.
A total of 13 hospitalized participants newly diagnosed with type 1 diabetes, aged 1~14 years old, were involved in the study.
Sensor glucose measurements on days 2, 3, 6, 7, 12, and 13 of wear were compared with venous BG.
During these days, the venous BG results were obtained either 4 or 7 times per day.
Results.
The accuracy was evaluated against venous BG, with 469 of 469 (100.0%) sensor and venous BG pairs within consensus error grid zones A and B, including 94.7% in zone A.
The overall mean absolute relative difference (MARD) was 11.67%.
The MARD of blood glucose lower than 4.0 mmol/L (MARD=16.89%) was higher than blood glucose between 4 and 10 mmol/L (MARD=11.58%) and blood glucose higher than 10 mmol/L (MARD=7.79%).
Compared to venous BG, the MARDs of wear days 2, 3, 6, 7, 12, and 13 were 11.53%, 9.66%, 11.79%, 10.89%, 13.18%, and 13.92%, respectively, with no statistically significant difference (P=0.25).
The median ARD was highest when the glucose decreased >0.11 mmol/L/min (20.27%) and lower than 10.00% when the glucose changed between 0.06 and 0.11 mmol/L/min, changed <0.06 mmol/L/min, and increased >0.11 mmol/L/min.
Conclusions.
The accuracy of the system is good and remains stable over 14 days of wear; however, the accuracy depends on the glucose level and rates of glucose concentration changes.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Cao, Bing-Yan& Wang, Rui& Gong, Chun-xiu& Wu, Di& Su, Chang& Chen, Jiajia…[et al.]. 2019. An Evaluation of the Accuracy of a Flash Glucose Monitoring System in Children with Diabetes in comparison with Venous Blood Glucose. Journal of Diabetes Research،Vol. 2019, no. 2019, pp.1-7.
https://search.emarefa.net/detail/BIM-1172966
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Cao, Bing-Yan…[et al.]. An Evaluation of the Accuracy of a Flash Glucose Monitoring System in Children with Diabetes in comparison with Venous Blood Glucose. Journal of Diabetes Research No. 2019 (2019), pp.1-7.
https://search.emarefa.net/detail/BIM-1172966
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Cao, Bing-Yan& Wang, Rui& Gong, Chun-xiu& Wu, Di& Su, Chang& Chen, Jiajia…[et al.]. An Evaluation of the Accuracy of a Flash Glucose Monitoring System in Children with Diabetes in comparison with Venous Blood Glucose. Journal of Diabetes Research. 2019. Vol. 2019, no. 2019, pp.1-7.
https://search.emarefa.net/detail/BIM-1172966
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-1172966
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر