An Evaluation of the Accuracy of a Flash Glucose Monitoring System in Children with Diabetes in comparison with Venous Blood Glucose

Joint Authors

Cao, Bing-Yan
Gong, Chun-xiu
Wu, Di
Liu, Min
Li, Wenjing
Su, Chang
Liang, Xuejun
Wang, Rui
Chen, Jiajia
Yi, Yajun

Source

Journal of Diabetes Research

Issue

Vol. 2019, Issue 2019 (31 Dec. 2019), pp.1-7, 7 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2019-09-09

Country of Publication

Egypt

No. of Pages

7

Main Subjects

Diseases
Medicine

Abstract EN

Aims.

To evaluate the performance of a factory-calibrated flash glucose monitoring system in children with diabetes compared to venous blood glucose (BG).

Methods.

A total of 13 hospitalized participants newly diagnosed with type 1 diabetes, aged 1~14 years old, were involved in the study.

Sensor glucose measurements on days 2, 3, 6, 7, 12, and 13 of wear were compared with venous BG.

During these days, the venous BG results were obtained either 4 or 7 times per day.

Results.

The accuracy was evaluated against venous BG, with 469 of 469 (100.0%) sensor and venous BG pairs within consensus error grid zones A and B, including 94.7% in zone A.

The overall mean absolute relative difference (MARD) was 11.67%.

The MARD of blood glucose lower than 4.0 mmol/L (MARD=16.89%) was higher than blood glucose between 4 and 10 mmol/L (MARD=11.58%) and blood glucose higher than 10 mmol/L (MARD=7.79%).

Compared to venous BG, the MARDs of wear days 2, 3, 6, 7, 12, and 13 were 11.53%, 9.66%, 11.79%, 10.89%, 13.18%, and 13.92%, respectively, with no statistically significant difference (P=0.25).

The median ARD was highest when the glucose decreased >0.11 mmol/L/min (20.27%) and lower than 10.00% when the glucose changed between 0.06 and 0.11 mmol/L/min, changed <0.06 mmol/L/min, and increased >0.11 mmol/L/min.

Conclusions.

The accuracy of the system is good and remains stable over 14 days of wear; however, the accuracy depends on the glucose level and rates of glucose concentration changes.

American Psychological Association (APA)

Cao, Bing-Yan& Wang, Rui& Gong, Chun-xiu& Wu, Di& Su, Chang& Chen, Jiajia…[et al.]. 2019. An Evaluation of the Accuracy of a Flash Glucose Monitoring System in Children with Diabetes in comparison with Venous Blood Glucose. Journal of Diabetes Research،Vol. 2019, no. 2019, pp.1-7.
https://search.emarefa.net/detail/BIM-1172966

Modern Language Association (MLA)

Cao, Bing-Yan…[et al.]. An Evaluation of the Accuracy of a Flash Glucose Monitoring System in Children with Diabetes in comparison with Venous Blood Glucose. Journal of Diabetes Research No. 2019 (2019), pp.1-7.
https://search.emarefa.net/detail/BIM-1172966

American Medical Association (AMA)

Cao, Bing-Yan& Wang, Rui& Gong, Chun-xiu& Wu, Di& Su, Chang& Chen, Jiajia…[et al.]. An Evaluation of the Accuracy of a Flash Glucose Monitoring System in Children with Diabetes in comparison with Venous Blood Glucose. Journal of Diabetes Research. 2019. Vol. 2019, no. 2019, pp.1-7.
https://search.emarefa.net/detail/BIM-1172966

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-1172966