Simultaneous Functional and Morphological Assessment of Left Atrial Appendage by 3D Virtual Models

المؤلفون المشاركون

Landini, Luigi
Berti, Sergio
Vivoli, Giorgia
Gasparotti, Emanuele
Rezzaghi, Marco
Cerone, Elisa
Mariani, Massimiliano
Celi, Simona
Positano, Vincenzo

المصدر

Journal of Healthcare Engineering

العدد

المجلد 2019، العدد 2019 (31 ديسمبر/كانون الأول 2019)، ص ص. 1-8، 8ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2019-05-21

دولة النشر

مصر

عدد الصفحات

8

التخصصات الرئيسية

الصحة العامة
الطب البشري

الملخص EN

Purpose.

The left atrial appendage (LAA) is responsible for thrombus formation in patients with atrial fibrillation.

The evaluation of both LAA function and morphology is crucial for the patient characterization and the preprocedural planning of LAA closure intervention.

Despite the availability of 3D imaging modalities, the current standard image analysis is based on manual delineation of the LAA contours on 2D views.

Methods.

In this study, a comprehensive approach based on a full 3D analysis of the tomographic dataset by surface extraction and processing (3D-S) is presented.

The proposed method allows extracting functional and morphologic information in the entire cardiac cycle by minimalizing manual user interaction.

The proposed methodology has been validated on ten computer tomography datasets.

Results.

The proposed 3D-S method was feasible in all cases.

Reproducibility was improved with respect to the reference 2D manual procedure (2D-S) (coefficient of variation 2.9 vs.

4.1% for diastolic ostium area; 3.8 vs.

6.1% for systolic ostium area; 2.4 vs.

5.3% for diastolic LAA volume; 2.7 vs.

5.9% for systolic LAA volume; and 7.7 vs.

17.1% for LAA ejection fraction).

No significant differences were found between 2D-S and 3D-S measurements.

Conclusions.

In this study, we introduced a fully 3D approach for LAA characterization, allowing the simultaneous assessment of LAA function and geometry.

The proposed approach could be used to improve the patient selection and the best sizing of the device for LAA closure and to allow a patient-specific 3D printing.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Vivoli, Giorgia& Gasparotti, Emanuele& Rezzaghi, Marco& Cerone, Elisa& Mariani, Massimiliano& Landini, Luigi…[et al.]. 2019. Simultaneous Functional and Morphological Assessment of Left Atrial Appendage by 3D Virtual Models. Journal of Healthcare Engineering،Vol. 2019, no. 2019, pp.1-8.
https://search.emarefa.net/detail/BIM-1175335

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Vivoli, Giorgia…[et al.]. Simultaneous Functional and Morphological Assessment of Left Atrial Appendage by 3D Virtual Models. Journal of Healthcare Engineering No. 2019 (2019), pp.1-8.
https://search.emarefa.net/detail/BIM-1175335

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Vivoli, Giorgia& Gasparotti, Emanuele& Rezzaghi, Marco& Cerone, Elisa& Mariani, Massimiliano& Landini, Luigi…[et al.]. Simultaneous Functional and Morphological Assessment of Left Atrial Appendage by 3D Virtual Models. Journal of Healthcare Engineering. 2019. Vol. 2019, no. 2019, pp.1-8.
https://search.emarefa.net/detail/BIM-1175335

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1175335