Evaluation of Tensile Force in a Porcine Trachea Using a Reflective Optical Method
المؤلفون المشاركون
Lomelí-Mejía, Pedro A.
Araujo-Monsalvo, Víctor M.
López-Sánchez, Alberto
Huerta-Azuara, Andrés
Aguirre-Palomé, Crystal
Domínguez-Hernández, Víctor M.
Cruz-Orea, Alfredo
المصدر
العدد
المجلد 2020، العدد 2020 (31 ديسمبر/كانون الأول 2020)، ص ص. 1-6، 6ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2020-03-21
دولة النشر
مصر
عدد الصفحات
6
التخصصات الرئيسية
الملخص EN
The trachea serves a variety of functions during breathing and swallowing.
In cases requiring a partial resection, the treatment may involve anastomosis of the healthy ends of the trachea.
The tension in the anastomosis should not exceed 1,000 g.
Traditional methods for estimating tension in the trachea are very invasive.
The aim of this work is to develop a noninvasive optical system devoted to estimate the tension of the trachea in vivo.
An optical system was designed using a light source and a photosensor.
To determine the most suitable wavelength for the light source, a photoacoustic spectroscopy study was performed.
To test the system, a cadaveric pig trachea was mounted on a universal testing machine, and it was subjected to three tensile tests of up to 30 mm elongation.
The optical response was measured at 0, 10, 20, and 30 mm elongation.
An exponential response was observed so that the optical voltage-response curves were adjusted, and three exponential equations were derived to correlate the voltage with the optical response.
We can conclude that the proposed optical system is able to noninvasively estimate the tension of a cadaveric porcine trachea.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Lomelí-Mejía, Pedro A.& Cruz-Orea, Alfredo& Araujo-Monsalvo, Víctor M.& López-Sánchez, Alberto& Huerta-Azuara, Andrés& Aguirre-Palomé, Crystal…[et al.]. 2020. Evaluation of Tensile Force in a Porcine Trachea Using a Reflective Optical Method. Journal of Spectroscopy،Vol. 2020, no. 2020, pp.1-6.
https://search.emarefa.net/detail/BIM-1190825
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Lomelí-Mejía, Pedro A.…[et al.]. Evaluation of Tensile Force in a Porcine Trachea Using a Reflective Optical Method. Journal of Spectroscopy No. 2020 (2020), pp.1-6.
https://search.emarefa.net/detail/BIM-1190825
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Lomelí-Mejía, Pedro A.& Cruz-Orea, Alfredo& Araujo-Monsalvo, Víctor M.& López-Sánchez, Alberto& Huerta-Azuara, Andrés& Aguirre-Palomé, Crystal…[et al.]. Evaluation of Tensile Force in a Porcine Trachea Using a Reflective Optical Method. Journal of Spectroscopy. 2020. Vol. 2020, no. 2020, pp.1-6.
https://search.emarefa.net/detail/BIM-1190825
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-1190825
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر