Evaluation of Body Composition in Hemodialysis Thai Patients: Comparison between Two Models of Bioelectrical Impedance Analyzer and Dual-Energy X-Ray Absorptiometry

المؤلفون المشاركون

Vallibhakara, Sakda Arj-Ong
Jayanama, Kulapong
Srisuwarn, Praopilad
Chattranukulchai Shantavasinkul, Prapimporn
Kantachuvesiri, Surasak
Komindr, Surat
Sritara, Chanika
Putadechakum, Supanee

المصدر

Journal of Nutrition and Metabolism

العدد

المجلد 2018، العدد 2018 (31 ديسمبر/كانون الأول 2018)، ص ص. 1-11، 11ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2018-08-05

دولة النشر

مصر

عدد الصفحات

11

التخصصات الرئيسية

التغذية والحميات

الملخص EN

Background.

Body composition measurement is very important for early nutritional care in hemodialysis patients.

Dual-energy X-ray absorptiometry (DXA) is a gold standard test, but clinically limited.

Bioelectrical impedance analysis (BIA) with multifrequency technique is a practical and reliable tool.

Objective.

This cross-sectional study was aimed to compare the agreement of BIA with DXA in measurement of body composition in hemodialysis patients and to evaluate their associated factors.

Methods.

Body composition was measured by 2 BIA methods (InBody S10 and InBody 720) and DXA after a hemodialysis session.

A total of 69 measurements were included.

Pearson’s correlation and Bland and Altman analysis were used to determine the correlation of body composition between methods and to compare the methods agreement, respectively.

Results.

The correlation coefficients of body compositions were strong between DXA and InBody S10 (fat mass index (FMI): r=0.95, fat-free mass index (FFMI): r=0.78) and also between DXA and InBody 720 (FMI: r=0.96, FFMI: r=0.81).

Comparing to DXA, the means of each body composition measured by InBody S10 method were not significantly different in each gender, but differences were found in FM, %FM, and FMI measured by InBody 720.

Conclusions.

In maintenance hemodialysis patients, the measurement of body composition with DXA and both BIA methods had highly significant correlations; practically, BIA method could be used as an instrument to follow FM and FFM and to measure the edematous stage.

Further studies with large populations are warranted.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Jayanama, Kulapong& Putadechakum, Supanee& Srisuwarn, Praopilad& Vallibhakara, Sakda Arj-Ong& Chattranukulchai Shantavasinkul, Prapimporn& Sritara, Chanika…[et al.]. 2018. Evaluation of Body Composition in Hemodialysis Thai Patients: Comparison between Two Models of Bioelectrical Impedance Analyzer and Dual-Energy X-Ray Absorptiometry. Journal of Nutrition and Metabolism،Vol. 2018, no. 2018, pp.1-11.
https://search.emarefa.net/detail/BIM-1195384

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Jayanama, Kulapong…[et al.]. Evaluation of Body Composition in Hemodialysis Thai Patients: Comparison between Two Models of Bioelectrical Impedance Analyzer and Dual-Energy X-Ray Absorptiometry. Journal of Nutrition and Metabolism No. 2018 (2018), pp.1-11.
https://search.emarefa.net/detail/BIM-1195384

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Jayanama, Kulapong& Putadechakum, Supanee& Srisuwarn, Praopilad& Vallibhakara, Sakda Arj-Ong& Chattranukulchai Shantavasinkul, Prapimporn& Sritara, Chanika…[et al.]. Evaluation of Body Composition in Hemodialysis Thai Patients: Comparison between Two Models of Bioelectrical Impedance Analyzer and Dual-Energy X-Ray Absorptiometry. Journal of Nutrition and Metabolism. 2018. Vol. 2018, no. 2018, pp.1-11.
https://search.emarefa.net/detail/BIM-1195384

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1195384