Verification of the Dose Reduction Effect via Diluted Injection in Dual-Energy Computed Tomography Using a Human Blood Flow Phantom

المؤلفون المشاركون

Tomita, Hironobu
Shibata, Koichi

المصدر

Radiology Research and Practice

العدد

المجلد 2019، العدد 2019 (31 ديسمبر/كانون الأول 2019)، ص ص. 1-9، 9ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2019-04-01

دولة النشر

مصر

عدد الصفحات

9

التخصصات الرئيسية

الطب البشري

الملخص EN

Purpose.

We sought to examine the possibility of reducing the contrast medium dosage in dual-energy imaging using a saline-mixed injection with a virtual monochromatic energy method of dual-source computed tomography (CT).

Methods.

An X-ray CT (SOMATOM Definition Flash: Siemens, Nurnberg, Germany) was employed.

The mixing ratio of contrast medium and saline was gradually changed by 10%, followed by a mixed injection into a dynamic blood flow phantom (Nemoto Kyorindo, Japan) which is a hemodynamic simulation phantom to obtain time-enhancement curves (TECs).

Exactly 64 TECs were prepared for each mixing ratio by changing the energy from 40 to 75 keV for monoenergetic imaging.

The relationship between the image standard deviation (SD) and the energy of the virtual monochromatic image was determined.

Combinations of the mixing ratio and energy (keV), which can maintain high CT numbers and low image SDs for 3D imaging, were tested, and the reduction rate of the contrast medium was calculated.

Results.

The TECs for the mixed injection method changed linearly with the dilution rates.

The mixing ratios were strongly correlated with the maximum CT number of the TEC (R2 = 0.98).

Contrast CT numbers and image SDs increased by approximately 20% and 25%, respectively, as the energy decreased by 5 keV.

The optimal conditions for reducing the contrast medium dose were a mixing ratio of 6:4 and 55 keV of energy.

Conclusion.

The virtual monochromatic energy method reduced the contrast medium dosage by up to 40% for three-dimensional CT-angio (3DCTA) tests.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Tomita, Hironobu& Shibata, Koichi. 2019. Verification of the Dose Reduction Effect via Diluted Injection in Dual-Energy Computed Tomography Using a Human Blood Flow Phantom. Radiology Research and Practice،Vol. 2019, no. 2019, pp.1-9.
https://search.emarefa.net/detail/BIM-1207551

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Tomita, Hironobu& Shibata, Koichi. Verification of the Dose Reduction Effect via Diluted Injection in Dual-Energy Computed Tomography Using a Human Blood Flow Phantom. Radiology Research and Practice No. 2019 (2019), pp.1-9.
https://search.emarefa.net/detail/BIM-1207551

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Tomita, Hironobu& Shibata, Koichi. Verification of the Dose Reduction Effect via Diluted Injection in Dual-Energy Computed Tomography Using a Human Blood Flow Phantom. Radiology Research and Practice. 2019. Vol. 2019, no. 2019, pp.1-9.
https://search.emarefa.net/detail/BIM-1207551

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1207551