Verification of the Dose Reduction Effect via Diluted Injection in Dual-Energy Computed Tomography Using a Human Blood Flow Phantom

Joint Authors

Tomita, Hironobu
Shibata, Koichi

Source

Radiology Research and Practice

Issue

Vol. 2019, Issue 2019 (31 Dec. 2019), pp.1-9, 9 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2019-04-01

Country of Publication

Egypt

No. of Pages

9

Main Subjects

Medicine

Abstract EN

Purpose.

We sought to examine the possibility of reducing the contrast medium dosage in dual-energy imaging using a saline-mixed injection with a virtual monochromatic energy method of dual-source computed tomography (CT).

Methods.

An X-ray CT (SOMATOM Definition Flash: Siemens, Nurnberg, Germany) was employed.

The mixing ratio of contrast medium and saline was gradually changed by 10%, followed by a mixed injection into a dynamic blood flow phantom (Nemoto Kyorindo, Japan) which is a hemodynamic simulation phantom to obtain time-enhancement curves (TECs).

Exactly 64 TECs were prepared for each mixing ratio by changing the energy from 40 to 75 keV for monoenergetic imaging.

The relationship between the image standard deviation (SD) and the energy of the virtual monochromatic image was determined.

Combinations of the mixing ratio and energy (keV), which can maintain high CT numbers and low image SDs for 3D imaging, were tested, and the reduction rate of the contrast medium was calculated.

Results.

The TECs for the mixed injection method changed linearly with the dilution rates.

The mixing ratios were strongly correlated with the maximum CT number of the TEC (R2 = 0.98).

Contrast CT numbers and image SDs increased by approximately 20% and 25%, respectively, as the energy decreased by 5 keV.

The optimal conditions for reducing the contrast medium dose were a mixing ratio of 6:4 and 55 keV of energy.

Conclusion.

The virtual monochromatic energy method reduced the contrast medium dosage by up to 40% for three-dimensional CT-angio (3DCTA) tests.

American Psychological Association (APA)

Tomita, Hironobu& Shibata, Koichi. 2019. Verification of the Dose Reduction Effect via Diluted Injection in Dual-Energy Computed Tomography Using a Human Blood Flow Phantom. Radiology Research and Practice،Vol. 2019, no. 2019, pp.1-9.
https://search.emarefa.net/detail/BIM-1207551

Modern Language Association (MLA)

Tomita, Hironobu& Shibata, Koichi. Verification of the Dose Reduction Effect via Diluted Injection in Dual-Energy Computed Tomography Using a Human Blood Flow Phantom. Radiology Research and Practice No. 2019 (2019), pp.1-9.
https://search.emarefa.net/detail/BIM-1207551

American Medical Association (AMA)

Tomita, Hironobu& Shibata, Koichi. Verification of the Dose Reduction Effect via Diluted Injection in Dual-Energy Computed Tomography Using a Human Blood Flow Phantom. Radiology Research and Practice. 2019. Vol. 2019, no. 2019, pp.1-9.
https://search.emarefa.net/detail/BIM-1207551

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-1207551