Forces Applied during Transvenous Implantable Cardioverter Defibrillator Lead Removal

المؤلفون المشاركون

Grebmer, Christian
Lennerz, Carsten
von Olshausen, Gesa
Ammar, Sonia
Buiatti, Alessandra
Pavaci, Herribert
Kolb, Christof
Semmler, Verena
Reents, Tilko
Deisenhofer, Isabel

المصدر

BioMed Research International

العدد

المجلد 2014، العدد 2014 (31 ديسمبر/كانون الأول 2014)، ص ص. 1-5، 5ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2014-05-21

دولة النشر

مصر

عدد الصفحات

5

التخصصات الرئيسية

الطب البشري

الملخص EN

Methods.

17 physicians, experienced in transvenous lead removal, performed a lead extraction manoeuvre of an ICD lead on a torso phantom.

They were advised to stop traction only when further traction would be considered as harmful to the patient or when—based on their experience—a change in the extraction strategy was indicated.

Traction forces were recorded with a digital precision gauge.

Results.

Median traction forces on the endocardium were 10.9 N (range from 3.0 N to 24.7 N and interquartile range from 7.9 to 15.3).

Forces applied to the proximal end were estimated to be 10% higher than those measured at the tip of the lead due to a friction loss.

Conclusion.

A traction force of around 11 N is typically exerted during standard transvenous extraction of ICD leads.

A traction threshold for a safe procedure derived from a pool of experienced extractionists may be helpful for the development of required adequate simulator trainings.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Lennerz, Carsten& Pavaci, Herribert& Grebmer, Christian& von Olshausen, Gesa& Semmler, Verena& Buiatti, Alessandra…[et al.]. 2014. Forces Applied during Transvenous Implantable Cardioverter Defibrillator Lead Removal. BioMed Research International،Vol. 2014, no. 2014, pp.1-5.
https://search.emarefa.net/detail/BIM-452597

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Lennerz, Carsten…[et al.]. Forces Applied during Transvenous Implantable Cardioverter Defibrillator Lead Removal. BioMed Research International No. 2014 (2014), pp.1-5.
https://search.emarefa.net/detail/BIM-452597

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Lennerz, Carsten& Pavaci, Herribert& Grebmer, Christian& von Olshausen, Gesa& Semmler, Verena& Buiatti, Alessandra…[et al.]. Forces Applied during Transvenous Implantable Cardioverter Defibrillator Lead Removal. BioMed Research International. 2014. Vol. 2014, no. 2014, pp.1-5.
https://search.emarefa.net/detail/BIM-452597

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-452597