Forces Applied during Transvenous Implantable Cardioverter Defibrillator Lead Removal

Joint Authors

Grebmer, Christian
Lennerz, Carsten
von Olshausen, Gesa
Ammar, Sonia
Buiatti, Alessandra
Pavaci, Herribert
Kolb, Christof
Semmler, Verena
Reents, Tilko
Deisenhofer, Isabel

Source

BioMed Research International

Issue

Vol. 2014, Issue 2014 (31 Dec. 2014), pp.1-5, 5 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2014-05-21

Country of Publication

Egypt

No. of Pages

5

Main Subjects

Medicine

Abstract EN

Methods.

17 physicians, experienced in transvenous lead removal, performed a lead extraction manoeuvre of an ICD lead on a torso phantom.

They were advised to stop traction only when further traction would be considered as harmful to the patient or when—based on their experience—a change in the extraction strategy was indicated.

Traction forces were recorded with a digital precision gauge.

Results.

Median traction forces on the endocardium were 10.9 N (range from 3.0 N to 24.7 N and interquartile range from 7.9 to 15.3).

Forces applied to the proximal end were estimated to be 10% higher than those measured at the tip of the lead due to a friction loss.

Conclusion.

A traction force of around 11 N is typically exerted during standard transvenous extraction of ICD leads.

A traction threshold for a safe procedure derived from a pool of experienced extractionists may be helpful for the development of required adequate simulator trainings.

American Psychological Association (APA)

Lennerz, Carsten& Pavaci, Herribert& Grebmer, Christian& von Olshausen, Gesa& Semmler, Verena& Buiatti, Alessandra…[et al.]. 2014. Forces Applied during Transvenous Implantable Cardioverter Defibrillator Lead Removal. BioMed Research International،Vol. 2014, no. 2014, pp.1-5.
https://search.emarefa.net/detail/BIM-452597

Modern Language Association (MLA)

Lennerz, Carsten…[et al.]. Forces Applied during Transvenous Implantable Cardioverter Defibrillator Lead Removal. BioMed Research International No. 2014 (2014), pp.1-5.
https://search.emarefa.net/detail/BIM-452597

American Medical Association (AMA)

Lennerz, Carsten& Pavaci, Herribert& Grebmer, Christian& von Olshausen, Gesa& Semmler, Verena& Buiatti, Alessandra…[et al.]. Forces Applied during Transvenous Implantable Cardioverter Defibrillator Lead Removal. BioMed Research International. 2014. Vol. 2014, no. 2014, pp.1-5.
https://search.emarefa.net/detail/BIM-452597

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-452597