Hemodynamic Surveillance of Ventricular Pacing Effectiveness with the Transvalvular Impedance Sensor

المؤلفون المشاركون

Felis, Salvatore
Di Gregorio, Franco
Virgilio, Alfredo
Barbetta, Alberto
Vasquez, Ludovico
Doria, Giuseppe
Calvi, Valeria
Pizzimenti, Giovanni
Lisi, Marco
Lisi, Francesco
Tempio, Donatella

المصدر

Advances in Medicine

العدد

المجلد 2014، العدد 2014 (31 ديسمبر/كانون الأول 2014)، ص ص. 1-7، 7ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2014-08-04

دولة النشر

مصر

عدد الصفحات

7

التخصصات الرئيسية

الطب البشري

الملخص EN

The Transvalvular Impedance (TVI) is derived between atrial and ventricular pacing electrodes.

A sharp TVI increase in systole is an ejection marker, allowing the hemodynamic surveillance of ventricular stimulation effectiveness in pacemaker patients.

At routine follow-up checks, the ventricular threshold test was managed by the stimulator with the supervision of a physician, who monitored the surface ECG.

When the energy scan resulted in capture loss, the TVI system must detect the failure and increase the output voltage.

A TVI signal suitable to this purpose was present in 85% of the tested patients.

A total of 230 capture failures, induced in 115 patients in both supine and sitting upright positions, were all promptly recognized by real-time TVI analysis (100% sensitivity).

The procedure was never interrupted by the physician, as the automatic energy regulation ensured full patient’s safety.

The pulse energy was then set at 4 times the threshold to test the alarm specificity during daily activity (sitting, standing up, and walking).

The median prevalence of false alarms was 0.336%.

The study shows that TVI-based ejection assessment is a valuable approach to the verification of pacing reliability and the autoregulation of ventricular stimulation energy.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Calvi, Valeria& Pizzimenti, Giovanni& Lisi, Marco& Doria, Giuseppe& Vasquez, Ludovico& Lisi, Francesco…[et al.]. 2014. Hemodynamic Surveillance of Ventricular Pacing Effectiveness with the Transvalvular Impedance Sensor. Advances in Medicine،Vol. 2014, no. 2014, pp.1-7.
https://search.emarefa.net/detail/BIM-462129

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Calvi, Valeria…[et al.]. Hemodynamic Surveillance of Ventricular Pacing Effectiveness with the Transvalvular Impedance Sensor. Advances in Medicine No. 2014 (2014), pp.1-7.
https://search.emarefa.net/detail/BIM-462129

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Calvi, Valeria& Pizzimenti, Giovanni& Lisi, Marco& Doria, Giuseppe& Vasquez, Ludovico& Lisi, Francesco…[et al.]. Hemodynamic Surveillance of Ventricular Pacing Effectiveness with the Transvalvular Impedance Sensor. Advances in Medicine. 2014. Vol. 2014, no. 2014, pp.1-7.
https://search.emarefa.net/detail/BIM-462129

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-462129