Hemodynamic Surveillance of Ventricular Pacing Effectiveness with the Transvalvular Impedance Sensor

Joint Authors

Felis, Salvatore
Di Gregorio, Franco
Virgilio, Alfredo
Barbetta, Alberto
Vasquez, Ludovico
Doria, Giuseppe
Calvi, Valeria
Pizzimenti, Giovanni
Lisi, Marco
Lisi, Francesco
Tempio, Donatella

Source

Advances in Medicine

Issue

Vol. 2014, Issue 2014 (31 Dec. 2014), pp.1-7, 7 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2014-08-04

Country of Publication

Egypt

No. of Pages

7

Main Subjects

Medicine

Abstract EN

The Transvalvular Impedance (TVI) is derived between atrial and ventricular pacing electrodes.

A sharp TVI increase in systole is an ejection marker, allowing the hemodynamic surveillance of ventricular stimulation effectiveness in pacemaker patients.

At routine follow-up checks, the ventricular threshold test was managed by the stimulator with the supervision of a physician, who monitored the surface ECG.

When the energy scan resulted in capture loss, the TVI system must detect the failure and increase the output voltage.

A TVI signal suitable to this purpose was present in 85% of the tested patients.

A total of 230 capture failures, induced in 115 patients in both supine and sitting upright positions, were all promptly recognized by real-time TVI analysis (100% sensitivity).

The procedure was never interrupted by the physician, as the automatic energy regulation ensured full patient’s safety.

The pulse energy was then set at 4 times the threshold to test the alarm specificity during daily activity (sitting, standing up, and walking).

The median prevalence of false alarms was 0.336%.

The study shows that TVI-based ejection assessment is a valuable approach to the verification of pacing reliability and the autoregulation of ventricular stimulation energy.

American Psychological Association (APA)

Calvi, Valeria& Pizzimenti, Giovanni& Lisi, Marco& Doria, Giuseppe& Vasquez, Ludovico& Lisi, Francesco…[et al.]. 2014. Hemodynamic Surveillance of Ventricular Pacing Effectiveness with the Transvalvular Impedance Sensor. Advances in Medicine،Vol. 2014, no. 2014, pp.1-7.
https://search.emarefa.net/detail/BIM-462129

Modern Language Association (MLA)

Calvi, Valeria…[et al.]. Hemodynamic Surveillance of Ventricular Pacing Effectiveness with the Transvalvular Impedance Sensor. Advances in Medicine No. 2014 (2014), pp.1-7.
https://search.emarefa.net/detail/BIM-462129

American Medical Association (AMA)

Calvi, Valeria& Pizzimenti, Giovanni& Lisi, Marco& Doria, Giuseppe& Vasquez, Ludovico& Lisi, Francesco…[et al.]. Hemodynamic Surveillance of Ventricular Pacing Effectiveness with the Transvalvular Impedance Sensor. Advances in Medicine. 2014. Vol. 2014, no. 2014, pp.1-7.
https://search.emarefa.net/detail/BIM-462129

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-462129