Establishment of a Swine Model for Validation of Perfusion Measurement by Dynamic Contrast-Enhanced Magnetic Resonance Imaging

المؤلفون المشاركون

Kramer, Martin
Hindel, Stefan
Wissmann, Andreas
Krüger, Christine
Sauerbrey, Anika
Nafz, Benno
Lüdemann, Lutz
Maaß, Marc

المصدر

BioMed Research International

العدد

المجلد 2014، العدد 2014 (31 ديسمبر/كانون الأول 2014)، ص ص. 1-9، 9ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2014-02-27

دولة النشر

مصر

عدد الصفحات

9

التخصصات الرئيسية

الطب البشري

الملخص EN

The aim of the study was to develop a suitable animal model for validating dynamic contrast-enhanced magnetic resonance imaging perfusion measurements.

A total of 8 pigs were investigated by DCE-MRI.

Perfusion was determined on the hind leg musculature.

An ultrasound flow probe placed around the femoral artery provided flow measurements independent of MRI and served as the standard of reference.

Images were acquired on a 1.5 T MRI scanner using a 3D T1-weighted gradient-echo sequence.

An arterial catheter for local injection was implanted in the femoral artery.

Continuous injection of adenosine for vasodilation resulted in steady blood flow levels up to four times the baseline level.

In this way, three different stable perfusion levels were induced and measured.

A central venous catheter was used for injection of two different types of contrast media.

A low-molecular weight contrast medium and a blood pool contrast medium were used.

A total of 6 perfusion measurements were performed with a time interval of about 20–25 min without significant differences in the arterial input functions.

In conclusion the accuracy of DCE-MRI-based perfusion measurement can be validated by comparison of the integrated perfusion signal of the hind leg musculature with the blood flow values measured with the ultrasound flow probe around the femoral artery.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Sauerbrey, Anika& Hindel, Stefan& Maaß, Marc& Krüger, Christine& Wissmann, Andreas& Kramer, Martin…[et al.]. 2014. Establishment of a Swine Model for Validation of Perfusion Measurement by Dynamic Contrast-Enhanced Magnetic Resonance Imaging. BioMed Research International،Vol. 2014, no. 2014, pp.1-9.
https://search.emarefa.net/detail/BIM-468325

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Sauerbrey, Anika…[et al.]. Establishment of a Swine Model for Validation of Perfusion Measurement by Dynamic Contrast-Enhanced Magnetic Resonance Imaging. BioMed Research International No. 2014 (2014), pp.1-9.
https://search.emarefa.net/detail/BIM-468325

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Sauerbrey, Anika& Hindel, Stefan& Maaß, Marc& Krüger, Christine& Wissmann, Andreas& Kramer, Martin…[et al.]. Establishment of a Swine Model for Validation of Perfusion Measurement by Dynamic Contrast-Enhanced Magnetic Resonance Imaging. BioMed Research International. 2014. Vol. 2014, no. 2014, pp.1-9.
https://search.emarefa.net/detail/BIM-468325

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-468325