Autoinflation Leading to Failure of Two-Piece Ambicor Implantable Penile Prosthesis : An Outcome from a Methodical Treatment of Recalcitrant Stuttering Priapism
المؤلفون المشاركون
Köhler, Tobias S.
West, Bradford L.
McVary, Kevin T.
Welliver, R. Charles
Fonseca, Adam N.
المصدر
العدد
المجلد 2014، العدد 2014 (31 ديسمبر/كانون الأول 2014)، ص ص. 1-3، 3ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2014-04-17
دولة النشر
مصر
عدد الصفحات
3
التخصصات الرئيسية
الملخص EN
Introduction.
We present the case of a patient who received a two-piece Ambicor penile prosthesis for idiopathic recurrent “stuttering” priapism refractory to other treatment options.
The patient returned unable to deflate the device due to an interesting anatomically induced mechanical failure.
Aims.
To describe the method and findings of this inflatable prosthesis failure.
Results.
Prosthesis failure occurred due to restrictive corporal diameter and the unique characteristics of fluid reservoir location in the two-piece inflatable prosthesis.
The patient was successfully converted to a semirigid prosthesis with resolution of the pain that was due to his prosthesis autoinflation.
Conclusion.
Stuttering priapism remains a challenging clinical problem.
Penile implantation is a reasonable long-term solution in a patient refractory to less invasive options.
In patients with fibrotic corpora, a malleable device should be considered (at least temporarily) if unable to dilate comfortably to 13 mm.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Welliver, R. Charles& Fonseca, Adam N.& West, Bradford L.& McVary, Kevin T.& Köhler, Tobias S.. 2014. Autoinflation Leading to Failure of Two-Piece Ambicor Implantable Penile Prosthesis : An Outcome from a Methodical Treatment of Recalcitrant Stuttering Priapism. Case Reports in Urology،Vol. 2014, no. 2014, pp.1-3.
https://search.emarefa.net/detail/BIM-478956
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Welliver, R. Charles…[et al.]. Autoinflation Leading to Failure of Two-Piece Ambicor Implantable Penile Prosthesis : An Outcome from a Methodical Treatment of Recalcitrant Stuttering Priapism. Case Reports in Urology No. 2014 (2014), pp.1-3.
https://search.emarefa.net/detail/BIM-478956
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Welliver, R. Charles& Fonseca, Adam N.& West, Bradford L.& McVary, Kevin T.& Köhler, Tobias S.. Autoinflation Leading to Failure of Two-Piece Ambicor Implantable Penile Prosthesis : An Outcome from a Methodical Treatment of Recalcitrant Stuttering Priapism. Case Reports in Urology. 2014. Vol. 2014, no. 2014, pp.1-3.
https://search.emarefa.net/detail/BIM-478956
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-478956
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
![](/images/ebook-kashef.png)
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر
![](/images/kashef-image.png)