Autoinflation Leading to Failure of Two-Piece Ambicor Implantable Penile Prosthesis : An Outcome from a Methodical Treatment of Recalcitrant Stuttering Priapism

Joint Authors

Köhler, Tobias S.
West, Bradford L.
McVary, Kevin T.
Welliver, R. Charles
Fonseca, Adam N.

Source

Case Reports in Urology

Issue

Vol. 2014, Issue 2014 (31 Dec. 2014), pp.1-3, 3 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2014-04-17

Country of Publication

Egypt

No. of Pages

3

Main Subjects

Medicine

Abstract EN

Introduction.

We present the case of a patient who received a two-piece Ambicor penile prosthesis for idiopathic recurrent “stuttering” priapism refractory to other treatment options.

The patient returned unable to deflate the device due to an interesting anatomically induced mechanical failure.

Aims.

To describe the method and findings of this inflatable prosthesis failure.

Results.

Prosthesis failure occurred due to restrictive corporal diameter and the unique characteristics of fluid reservoir location in the two-piece inflatable prosthesis.

The patient was successfully converted to a semirigid prosthesis with resolution of the pain that was due to his prosthesis autoinflation.

Conclusion.

Stuttering priapism remains a challenging clinical problem.

Penile implantation is a reasonable long-term solution in a patient refractory to less invasive options.

In patients with fibrotic corpora, a malleable device should be considered (at least temporarily) if unable to dilate comfortably to 13 mm.

American Psychological Association (APA)

Welliver, R. Charles& Fonseca, Adam N.& West, Bradford L.& McVary, Kevin T.& Köhler, Tobias S.. 2014. Autoinflation Leading to Failure of Two-Piece Ambicor Implantable Penile Prosthesis : An Outcome from a Methodical Treatment of Recalcitrant Stuttering Priapism. Case Reports in Urology،Vol. 2014, no. 2014, pp.1-3.
https://search.emarefa.net/detail/BIM-478956

Modern Language Association (MLA)

Welliver, R. Charles…[et al.]. Autoinflation Leading to Failure of Two-Piece Ambicor Implantable Penile Prosthesis : An Outcome from a Methodical Treatment of Recalcitrant Stuttering Priapism. Case Reports in Urology No. 2014 (2014), pp.1-3.
https://search.emarefa.net/detail/BIM-478956

American Medical Association (AMA)

Welliver, R. Charles& Fonseca, Adam N.& West, Bradford L.& McVary, Kevin T.& Köhler, Tobias S.. Autoinflation Leading to Failure of Two-Piece Ambicor Implantable Penile Prosthesis : An Outcome from a Methodical Treatment of Recalcitrant Stuttering Priapism. Case Reports in Urology. 2014. Vol. 2014, no. 2014, pp.1-3.
https://search.emarefa.net/detail/BIM-478956

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-478956