Potentiometric Determination of Ketotifen Fumarate in Pharmaceutical Preparations and Urine Using Carbon Paste and PVC Membrane Selective Electrodes

المؤلفون المشاركون

Hwehy, Mohammad M. A.
Frag, Eman Y. Z.
Mohamed, Gehad G.
Khalil, Mohamed M.

المصدر

International Journal of Analytical Chemistry

العدد

المجلد 2011، العدد 2011 (31 ديسمبر/كانون الأول 2011)، ص ص. 1-7، 7ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2011-10-13

دولة النشر

مصر

عدد الصفحات

7

التخصصات الرئيسية

الكيمياء
علوم

الملخص EN

This study compares between unmodified carbon paste (CPE; the paste has no ion pair) and polyvinyl chloride (PVC) membrane selective electrodes that were used in potentiometric determination of ketotifen fumarate (KTF), where sodium tetraphenylborate (NaTPB) was used as titrant.

The performance characteristics of these sensors were evaluated according to IUPAC recommendations which reveal a fast, stable, and linear response for KTF over the concentration range of 10-7 to 10-2 mol L-1.

The electrodes show Nernstian slope value of 52.51±0.20 and 51.51±0.25 mV decade-1 for CPE and PVC membrane electrodes at 30∘C, respectively.

The potential is nearly stable over the pH range 3.0–6.0 and 2.0–7.0 for CPE and PVC membrane electrodes, respectively.

Selectivity coefficient values towards different inorganic cations, sugars, and amino acids reflect high selectivity of the prepared electrodes.

The electrodes responses at different temperatures were also studied, and long operational lifetime of 12 and 5 weeks for CPE and PVC membrane electrodes, respectively, were found.

These are used for determination of ketotifen fumarate using potentiometric titration, calibration, and standard addition methods in pure samples, its pharmaceutical preparations (Zaditen tablets), and biological fluid (urine).

The direct potentiometric determination of KTF using the proposed sensors gave recoveries % of 98.97±0.53 and 98.62±0.74 with RSD 1.42 and 0.63% for CPE and PVC membrane selective electrodes, respectively.

Validation of the method shows suitability of the proposed sensors for use in quality control assessment of KTF.

The obtained results were in a good agreement with those obtained using the reported spectrophotometric method.Erratum of “Potentiometric Determination of Ketotifen Fumarate in Pharmaceutical Preparations and Urine Using Carbon Paste and PVC Membrane Selective Electrodes”dx.doi.org/10.1155/2012/154907

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Frag, Eman Y. Z.& Mohamed, Gehad G.& Khalil, Mohamed M.& Hwehy, Mohammad M. A.. 2011. Potentiometric Determination of Ketotifen Fumarate in Pharmaceutical Preparations and Urine Using Carbon Paste and PVC Membrane Selective Electrodes. International Journal of Analytical Chemistry،Vol. 2011, no. 2011, pp.1-7.
https://search.emarefa.net/detail/BIM-484421

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Frag, Eman Y. Z.…[et al.]. Potentiometric Determination of Ketotifen Fumarate in Pharmaceutical Preparations and Urine Using Carbon Paste and PVC Membrane Selective Electrodes. International Journal of Analytical Chemistry No. 2011 (2011), pp.1-7.
https://search.emarefa.net/detail/BIM-484421

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Frag, Eman Y. Z.& Mohamed, Gehad G.& Khalil, Mohamed M.& Hwehy, Mohammad M. A.. Potentiometric Determination of Ketotifen Fumarate in Pharmaceutical Preparations and Urine Using Carbon Paste and PVC Membrane Selective Electrodes. International Journal of Analytical Chemistry. 2011. Vol. 2011, no. 2011, pp.1-7.
https://search.emarefa.net/detail/BIM-484421

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-484421