![](/images/graphics-bg.png)
Comparison of the Tendon Damage Caused by Four Different Anchor Systems Used in Transtendon Rotator Cuff Repair
المؤلفون المشاركون
O'Brien, Michael J.
Zhang, Qing-Song
Zhang, Qiuyang
Ge, Dongxia
You, Zongbing
Liu, Sen
Xue, Yun
Savoie, Felix H.
المصدر
العدد
المجلد 2012، العدد 2012 (31 ديسمبر/كانون الأول 2012)، ص ص. 1-6، 6ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2012-07-01
دولة النشر
مصر
عدد الصفحات
6
التخصصات الرئيسية
الملخص EN
Objectives.
The objective of this study was to compare the damage to the rotator cuff tendons caused by four different anchor systems.
Methods.
20 cadaveric human shoulder joints were used for transtendon insertion of four anchor systems.
The Healix Peek, Fastin RC, Bio-Corkscrew Suture, and Healix Transtend anchors were inserted through the tendons using standard transtendon procedures.
The areas of tendon damage were measured.
Results.
The areas of tendon damage (mean ± standard deviation, n=7) were 29.1 ± 4.3 mm2 for the Healix Peek anchor, 20.4 ± 2.3 mm2 for the Fastin RC anchor, 23.4 ± 1.2 mm2 for the Bio-Corkscrew Suture anchor, 13.7 ± 3.2 mm2 for the Healix Transtend anchor inserted directly, and 9.1 ± 2.1 mm2 for the Healix Transtend anchor inserted through the Percannula system (P<0.001 or P<0.001, compared to other anchors).
Conclusions.
In a cadaver transtendon rotator cuff repair model, smaller anchors caused less damage to the tendon tissues.
The Healix Transtend implant system caused the least damage to the tendon tissues.
Our findings suggest that smaller anchors should be considered when performing transtendon procedures to repair partial rotator cuff tears.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Zhang, Qing-Song& Liu, Sen& Zhang, Qiuyang& Xue, Yun& Ge, Dongxia& O'Brien, Michael J.…[et al.]. 2012. Comparison of the Tendon Damage Caused by Four Different Anchor Systems Used in Transtendon Rotator Cuff Repair. Advances in Orthopedics،Vol. 2012, no. 2012, pp.1-6.
https://search.emarefa.net/detail/BIM-499044
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Zhang, Qing-Song…[et al.]. Comparison of the Tendon Damage Caused by Four Different Anchor Systems Used in Transtendon Rotator Cuff Repair. Advances in Orthopedics No. 2012 (2012), pp.1-6.
https://search.emarefa.net/detail/BIM-499044
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Zhang, Qing-Song& Liu, Sen& Zhang, Qiuyang& Xue, Yun& Ge, Dongxia& O'Brien, Michael J.…[et al.]. Comparison of the Tendon Damage Caused by Four Different Anchor Systems Used in Transtendon Rotator Cuff Repair. Advances in Orthopedics. 2012. Vol. 2012, no. 2012, pp.1-6.
https://search.emarefa.net/detail/BIM-499044
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-499044
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
![](/images/ebook-kashef.png)
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر
![](/images/kashef-image.png)