The Use of a Modular Titanium Baseplate with a Press-Fit Keel Implanted with a Surface Cementing Technique for Primary Total Knee Arthroplasty

المؤلفون المشاركون

Erickson, Jill
Peters, Christopher L.
Evans, David
Christensen, Bryt A.
Severson, Erik P.
Pelt, Christopher E.
Widmer, Benjamin

المصدر

BioMed Research International

العدد

المجلد 2014، العدد 2014 (31 ديسمبر/كانون الأول 2014)، ص ص. 1-6، 6ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2014-05-14

دولة النشر

مصر

عدد الصفحات

6

التخصصات الرئيسية

الطب البشري

الملخص EN

Little data exists regarding outcomes following TKA performed with surface-cementation for the fixation of modular tibial baseplates with press-fit keels.

Thus, we retrospectively reviewed the clinical and radiographic outcomes of 439 consecutive primary TKAs performed with surface cemented tibial components.

There were 290 female patients and 149 male patients with average age of 62 years (range 30–84).

Two tibial components were revised for aseptic loosening (0.5%) and four tibial components (0.9%) were removed to improve instability (n=2) or malalignment (n=2).

Complications included 13 deep infections treated with 2-stage revision (12) and fusion (1).

These results support the surface cement technique with a modular grit-blasted titanium surface and cruciform stem during primary TKA.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Pelt, Christopher E.& Erickson, Jill& Christensen, Bryt A.& Widmer, Benjamin& Severson, Erik P.& Evans, David…[et al.]. 2014. The Use of a Modular Titanium Baseplate with a Press-Fit Keel Implanted with a Surface Cementing Technique for Primary Total Knee Arthroplasty. BioMed Research International،Vol. 2014, no. 2014, pp.1-6.
https://search.emarefa.net/detail/BIM-512572

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Pelt, Christopher E.…[et al.]. The Use of a Modular Titanium Baseplate with a Press-Fit Keel Implanted with a Surface Cementing Technique for Primary Total Knee Arthroplasty. BioMed Research International No. 2014 (2014), pp.1-6.
https://search.emarefa.net/detail/BIM-512572

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Pelt, Christopher E.& Erickson, Jill& Christensen, Bryt A.& Widmer, Benjamin& Severson, Erik P.& Evans, David…[et al.]. The Use of a Modular Titanium Baseplate with a Press-Fit Keel Implanted with a Surface Cementing Technique for Primary Total Knee Arthroplasty. BioMed Research International. 2014. Vol. 2014, no. 2014, pp.1-6.
https://search.emarefa.net/detail/BIM-512572

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-512572