Calculations of effective dose in diagnostic radiology

المؤلفون المشاركون

Habbani, F. I.
Sulayman, I. I.

المصدر

Sudan Journal of Basic Sciences. |n Series : Mathematical Science

العدد

المجلد 2012، العدد 17 (31 ديسمبر/كانون الأول 2012)، ص ص. 121-134، 14ص.

الناشر

معهد السودان للعلوم الطبيعية وحدة التأليف و النشر

تاريخ النشر

2012-12-31

دولة النشر

السودان

عدد الصفحات

14

التخصصات الرئيسية

الرياضيات

الملخص EN

The study was performed to compare two methods used for effective dose (ED) calculations in diagnostic radiology and to evaluate the ED distribution in radiological departments in four major hospitals in Khartoum, Sudan.

Dosimetric data for six radiographic projections and a sample of 346 patients were used in the study.

First, ED was calculated from entrance skin dose using NRPB-SR262 Monte Carlo data.

Second, energy imparted () to each patient was estimated using entrance skin exposure, half value layer and the beam area at the entrance plane.

ED was then estimated using projection specific ED/ conversion factors.

Mean ED values calculated using the two methods were: 21.3-23.4, 14.1-12.8, 7.9-8.5, 232-226, 215-223 and 91-85.6 μSv for chest PA, Skull AP/PA, Skull LAT, Pelvis AP, Lumbar Spine AP and Lumbar Spine LAT examinations, respectively.

The values obtained are in good agreement between themselves and with data reported in the literature

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Habbani, F. I.& Sulayman, I. I.. 2012. Calculations of effective dose in diagnostic radiology. Sudan Journal of Basic Sciences. |n Series : Mathematical Science،Vol. 2012, no. 17, pp.121-134.
https://search.emarefa.net/detail/BIM-754407

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Habbani, F. I.& Sulayman, I. I.. Calculations of effective dose in diagnostic radiology. Sudan Journal of Basic Sciences. |n Series : Mathematical Science No. 17 (2012), pp.121-134.
https://search.emarefa.net/detail/BIM-754407

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Habbani, F. I.& Sulayman, I. I.. Calculations of effective dose in diagnostic radiology. Sudan Journal of Basic Sciences. |n Series : Mathematical Science. 2012. Vol. 2012, no. 17, pp.121-134.
https://search.emarefa.net/detail/BIM-754407

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references : p. 132-134

رقم السجل

BIM-754407