Robust high resolution fat-water separation in the abdomen during free-breathing by self-gated 2D Radial TrueFISP imaging
المؤلفون المشاركون
Ababinah, Riyad S.
Benkert, Thomas
Breuer, Felix A.
المصدر
العدد
المجلد 9، العدد 2 (31 ديسمبر/كانون الأول 2016)، ص ص. 103-108، 6ص.
الناشر
جامعة اليرموك عمادة البحث العلمي و الدراسات العليا
تاريخ النشر
2016-12-31
دولة النشر
الأردن
عدد الصفحات
6
التخصصات الرئيسية
الملخص EN
Accurate high resolution fat-water separation in the abdomen is challenging due to respiratory motion.
In this work, we propose a robust high resolution fat-water separation strategy in the abdomen during free breathing by employing radial sampling with a golden angle increment.
To this end, a radial TrueFISP sequence was modified enabling the echo time TE to change from projection to projection, to force fat signals to behave in a conspicuous manner through time, so that they can be detected and separated from water signals through temporal processing.
Thus, the center signal (DC) of each radial readout can then be used for respiratory gating, allowing the generation of multiple images at different TEs at multiple breathing states.
Finally, any fat-water separation technique can be used to synthesize high resolution fat and water images at arbitrary breathing states.
Good separation of fat and water signals was achieved using the radial TrueFISP sequence during free breathing without streaking artifacts or blurring due to respiratory motion.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Ababinah, Riyad S.& Benkert, Thomas& Breuer, Felix A.. 2016. Robust high resolution fat-water separation in the abdomen during free-breathing by self-gated 2D Radial TrueFISP imaging. Jordan Journal of Physics،Vol. 9, no. 2, pp.103-108.
https://search.emarefa.net/detail/BIM-818887
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Ababinah, Riyad S.…[et al.]. Robust high resolution fat-water separation in the abdomen during free-breathing by self-gated 2D Radial TrueFISP imaging. Jordan Journal of Physics Vol. 9, no. 2 (2016), pp.103-108.
https://search.emarefa.net/detail/BIM-818887
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Ababinah, Riyad S.& Benkert, Thomas& Breuer, Felix A.. Robust high resolution fat-water separation in the abdomen during free-breathing by self-gated 2D Radial TrueFISP imaging. Jordan Journal of Physics. 2016. Vol. 9, no. 2, pp.103-108.
https://search.emarefa.net/detail/BIM-818887
نوع البيانات
مقالات
لغة النص
الإنجليزية
رقم السجل
BIM-818887
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر