Effect of Low Tube Voltage on Image Quality, Radiation Dose, and Low-Contrast Detectability at Abdominal Multidetector CT: Phantom Study
المؤلفون المشاركون
Lin, Jie
Tang, Kun
Li, Rui
Wang, Ling
Cao, Guoquan
Zheng, Xiangwu
المصدر
العدد
المجلد 2012، العدد 2012 (31 ديسمبر/كانون الأول 2012)، ص ص. 1-6، 6ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2012-04-23
دولة النشر
مصر
عدد الصفحات
6
التخصصات الرئيسية
الملخص EN
Purpose.
To investigate the effect of low tube voltage (80 kV) on image quality, radiation dose, and low-contrast detectability (LCD) at abdominal computed tomography (CT).
Materials and Methods.
A phantom containing low-contrast objects was scanned with a CT scanner at 80 and 120 kV, with tube current-time product settings at 150–650 mAs.
The differences between image noise, contrast-to-noise ratio (CNR), and scores of LCD obtained with 80 kV at 150–650 mAs and those obtained with 120 kV at 300 mAs were compared respectively.
Results.
The image noise substantially increased with low tube voltage.
However, with identical dose, use of 80 kV resulted in higher CNR compared with CNR at 120 kV.
There were no statistically significant difference in CNR and scores of LCD between 120 kV at 300 mAs and 80 kV at 550–650 mAs (P>0.05).
The relative dose delivered at 80 kV ranged from 58% at 550 mAs to 68% at 650 mAs.
Conclusion.
With a reduction of the tube voltage from 120 kV to 80 kV at abdominal CT, the radiation dose can be reduced by 32% to 42% without degradation of CNR and LCD.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Tang, Kun& Wang, Ling& Li, Rui& Lin, Jie& Zheng, Xiangwu& Cao, Guoquan. 2012. Effect of Low Tube Voltage on Image Quality, Radiation Dose, and Low-Contrast Detectability at Abdominal Multidetector CT: Phantom Study. BioMed Research International،Vol. 2012, no. 2012, pp.1-6.
https://search.emarefa.net/detail/BIM-991458
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Tang, Kun…[et al.]. Effect of Low Tube Voltage on Image Quality, Radiation Dose, and Low-Contrast Detectability at Abdominal Multidetector CT: Phantom Study. BioMed Research International No. 2012 (2012), pp.1-6.
https://search.emarefa.net/detail/BIM-991458
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Tang, Kun& Wang, Ling& Li, Rui& Lin, Jie& Zheng, Xiangwu& Cao, Guoquan. Effect of Low Tube Voltage on Image Quality, Radiation Dose, and Low-Contrast Detectability at Abdominal Multidetector CT: Phantom Study. BioMed Research International. 2012. Vol. 2012, no. 2012, pp.1-6.
https://search.emarefa.net/detail/BIM-991458
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-991458
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر