![](/images/graphics-bg.png)
Comparison of Pentax-AWS Airwayscope and Glidescope for Infant Tracheal Intubation by Anesthesiologists during Cardiopulmonary Arrest Simulation: A Randomized Crossover Trial
Joint Authors
Fujiwara, Shunsuke
Komasawa, Nobuyasu
Matsunami, Sayuri
Okada, Daisuke
Minami, Toshiaki
Source
Issue
Vol. 2015, Issue 2015 (31 Dec. 2015), pp.1-5, 5 p.
Publisher
Hindawi Publishing Corporation
Publication Date
2015-06-16
Country of Publication
Egypt
No. of Pages
5
Main Subjects
Abstract EN
Background.
Recent guidelines for infant cardiopulmonary resuscitation emphasize that all rescuers should minimize interruption of chest compressions, even for endotracheal intubation.
We compared the utility of the Pentax-AWS Airwayscope (AWS) with the Glidescope (GS) during chest compressions on an infant manikin.
Methods.
Twenty-four anesthesiologists with more than two years of experience performed tracheal intubation on an infant manikin using the AWS and GS, with or without chest compressions.
Results.
In GS trials, none of the participants failed without compressions, while three failed with compressions.
In AWS trials, all participants succeeded regardless of chest compressions.
Intubation time was significantly longer with chest compressions with the GS (P<0.05), but not with the AWS.
Difficulty of operation on a visual analog scale (VAS) for laryngoscopy did not increase significantly with chest compressions with either the GS or the AWS, while the VAS for tube passage through the glottis increased with compressions with the GS, but not with the AWS.
Conclusion.
We conclude that in infant simulations managed by anesthesiologists, the AWS performed better than the GS for endotracheal intubation with chest compressions.
American Psychological Association (APA)
Fujiwara, Shunsuke& Komasawa, Nobuyasu& Matsunami, Sayuri& Okada, Daisuke& Minami, Toshiaki. 2015. Comparison of Pentax-AWS Airwayscope and Glidescope for Infant Tracheal Intubation by Anesthesiologists during Cardiopulmonary Arrest Simulation: A Randomized Crossover Trial. BioMed Research International،Vol. 2015, no. 2015, pp.1-5.
https://search.emarefa.net/detail/BIM-1054562
Modern Language Association (MLA)
Fujiwara, Shunsuke…[et al.]. Comparison of Pentax-AWS Airwayscope and Glidescope for Infant Tracheal Intubation by Anesthesiologists during Cardiopulmonary Arrest Simulation: A Randomized Crossover Trial. BioMed Research International No. 2015 (2015), pp.1-5.
https://search.emarefa.net/detail/BIM-1054562
American Medical Association (AMA)
Fujiwara, Shunsuke& Komasawa, Nobuyasu& Matsunami, Sayuri& Okada, Daisuke& Minami, Toshiaki. Comparison of Pentax-AWS Airwayscope and Glidescope for Infant Tracheal Intubation by Anesthesiologists during Cardiopulmonary Arrest Simulation: A Randomized Crossover Trial. BioMed Research International. 2015. Vol. 2015, no. 2015, pp.1-5.
https://search.emarefa.net/detail/BIM-1054562
Data Type
Journal Articles
Language
English
Notes
Includes bibliographical references
Record ID