Hemodynamic Study of Flow Remodeling Stent Graft for the Treatment of Highly Angulated Abdominal Aortic Aneurysm

Joint Authors

Leo, Hwa-Liang
Yeow, Siang Lin

Source

Computational and Mathematical Methods in Medicine

Issue

Vol. 2016, Issue 2016 (31 Dec. 2016), pp.1-10, 10 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2016-05-09

Country of Publication

Egypt

No. of Pages

10

Main Subjects

Medicine

Abstract EN

This study investigates the effect of a novel flow remodeling stent graft (FRSG) on the hemodynamic characteristics in highly angulated abdominal aortic aneurysm based on computational fluid dynamics (CFD) approach.

An idealized aortic aneurysm with varying aortic neck angulations was constructed and CFD simulations were performed on nonstented models and stented models with FRSG.

The influence of FRSG intervention on the hemodynamic performance is analyzed and compared in terms of flow patterns, wall shear stress (WSS), and pressure distribution in the aneurysm.

The findings showed that aortic neck angulations significantly influence the velocity flow field in nonstented models, with larger angulations shifting the mainstream blood flow towards the center of the aorta.

By introducing FRSG treatment into the aneurysm, erratic flow recirculation pattern in the aneurysm sac diminishes while the average velocity magnitude in the aneurysm sac was reduced in the range of 39% to 53%.

FRSG intervention protects the aneurysm against the impacts of high velocity concentrated flow and decreases wall shear stress by more than 50%.

The simulation results highlighted that FRSG may effectively treat aneurysm with high aortic neck angulations via the mechanism of promoting thrombus formation and subsequently led to the resorption of the aneurysm.

American Psychological Association (APA)

Yeow, Siang Lin& Leo, Hwa-Liang. 2016. Hemodynamic Study of Flow Remodeling Stent Graft for the Treatment of Highly Angulated Abdominal Aortic Aneurysm. Computational and Mathematical Methods in Medicine،Vol. 2016, no. 2016, pp.1-10.
https://search.emarefa.net/detail/BIM-1100120

Modern Language Association (MLA)

Yeow, Siang Lin& Leo, Hwa-Liang. Hemodynamic Study of Flow Remodeling Stent Graft for the Treatment of Highly Angulated Abdominal Aortic Aneurysm. Computational and Mathematical Methods in Medicine No. 2016 (2016), pp.1-10.
https://search.emarefa.net/detail/BIM-1100120

American Medical Association (AMA)

Yeow, Siang Lin& Leo, Hwa-Liang. Hemodynamic Study of Flow Remodeling Stent Graft for the Treatment of Highly Angulated Abdominal Aortic Aneurysm. Computational and Mathematical Methods in Medicine. 2016. Vol. 2016, no. 2016, pp.1-10.
https://search.emarefa.net/detail/BIM-1100120

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-1100120