Accuracy of Bone Measurements in the Vicinity of Titanium Implants in CBCT Data Sets: A Comparison of Radiological and Histological Findings in Minipigs

Joint Authors

Hanken, Henning
Eichhorn, Wolfgang
Semmusch, Jan
Jung, Ole
Schön, Gerhard MSc
Schöllchen, Maximilian
Hahn, Michael
Stein, Jamal M.
Reitmeier, Aline
Precht, Clarissa
Smeets, Ralf
Gröbe, Alexander
Heiland, Max

Source

BioMed Research International

Issue

Vol. 2017, Issue 2017 (31 Dec. 2017), pp.1-9, 9 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2017-07-17

Country of Publication

Egypt

No. of Pages

9

Main Subjects

Medicine

Abstract EN

Purpose.

The aim of this animal study was the determination of accuracy of bone measurements in CBCT (cone-beam computed tomography) in close proximity to titanium implants.

Material and Methods.

Titanium implants were inserted in eight Göttingen minipigs.

60 implants were evaluated histologically in ground section specimen and radiologically in CBCT in regard to thickness of the buccal bone.

With random intercept models, the difference of histologic measurements and CBCT measurements of bone thickness was calculated.

Results.

The mean histological thickness of the buccal bone was 5.09 mm (CI 4.11–6.08 mm).

The four raters measured slightly less bone in CBCT than it was found in histology.

The random effect was not significant (p value 1.000).

Therefore, the Intraclass Correlation Coefficient (ICC) was 98.65% (CI 100.00–96.99%).

Conclusion.

CBCT is an accurate technique to measure even thin bone structures in the vicinity of titanium implants.

American Psychological Association (APA)

Gröbe, Alexander& Semmusch, Jan& Schöllchen, Maximilian& Hanken, Henning& Hahn, Michael& Eichhorn, Wolfgang…[et al.]. 2017. Accuracy of Bone Measurements in the Vicinity of Titanium Implants in CBCT Data Sets: A Comparison of Radiological and Histological Findings in Minipigs. BioMed Research International،Vol. 2017, no. 2017, pp.1-9.
https://search.emarefa.net/detail/BIM-1136456

Modern Language Association (MLA)

Gröbe, Alexander…[et al.]. Accuracy of Bone Measurements in the Vicinity of Titanium Implants in CBCT Data Sets: A Comparison of Radiological and Histological Findings in Minipigs. BioMed Research International No. 2017 (2017), pp.1-9.
https://search.emarefa.net/detail/BIM-1136456

American Medical Association (AMA)

Gröbe, Alexander& Semmusch, Jan& Schöllchen, Maximilian& Hanken, Henning& Hahn, Michael& Eichhorn, Wolfgang…[et al.]. Accuracy of Bone Measurements in the Vicinity of Titanium Implants in CBCT Data Sets: A Comparison of Radiological and Histological Findings in Minipigs. BioMed Research International. 2017. Vol. 2017, no. 2017, pp.1-9.
https://search.emarefa.net/detail/BIM-1136456

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID