An Evaluation of the Gap Sizes of 3-Unit Fixed Dental Prostheses Milled from Sintering Metal Blocks
Author
Source
Issue
Vol. 2017, Issue 2017 (31 Dec. 2017), pp.1-8, 8 p.
Publisher
Hindawi Publishing Corporation
Publication Date
2017-01-29
Country of Publication
Egypt
No. of Pages
8
Main Subjects
Abstract EN
This study assessed the clinical acceptability of sintering metal-fabricated 3-unit fixed dental prostheses (FDPs) based on gap sizes.
Ten specimens were prepared on research models by milling sintering metal blocks or by the lost-wax technique (LWC group).
Gap sizes were assessed at 12 points per abutment (premolar and molar), 24 points per specimen (480 points in a total in 20 specimens).
The measured points were categorized as marginal, axial wall, and occlusal for assessment in a silicone replica.
The silicone replica was cut through the mesiodistal and buccolingual center.
The four sections were magnified at 160x, and the thickness of the light body silicone was measured to determine the gap size, and gap size means were compared.
For the premolar part, the mean (standard deviation) gap size was nonsignificantly (p=0.139) smaller in the SMB group (68.6 ± 35.6 μm) than in the LWC group (69.6 ± 16.9 μm).
The mean molar gap was nonsignificantly smaller (p=0.852) in the LWC (73.9 ± 25.6 μm) than in the SMB (78.1 ± 37.4 μm) group.
The gap sizes were similar between the two groups.
Because the gap sizes were within the previously proposed clinically accepted limit, FDPs prepared by sintered metal block milling are clinically acceptable.
American Psychological Association (APA)
Jung, Jae-Kwan. 2017. An Evaluation of the Gap Sizes of 3-Unit Fixed Dental Prostheses Milled from Sintering Metal Blocks. BioMed Research International،Vol. 2017, no. 2017, pp.1-8.
https://search.emarefa.net/detail/BIM-1138743
Modern Language Association (MLA)
Jung, Jae-Kwan. An Evaluation of the Gap Sizes of 3-Unit Fixed Dental Prostheses Milled from Sintering Metal Blocks. BioMed Research International No. 2017 (2017), pp.1-8.
https://search.emarefa.net/detail/BIM-1138743
American Medical Association (AMA)
Jung, Jae-Kwan. An Evaluation of the Gap Sizes of 3-Unit Fixed Dental Prostheses Milled from Sintering Metal Blocks. BioMed Research International. 2017. Vol. 2017, no. 2017, pp.1-8.
https://search.emarefa.net/detail/BIM-1138743
Data Type
Journal Articles
Language
English
Notes
Includes bibliographical references
Record ID
BIM-1138743