Real-Time Chest Compression Quality Measurements by Smartphone Camera

Joint Authors

Engan, Kjersti
Meinich-Bache, Øyvind
Birkenes, Tonje Søraas
Myklebust, Helge

Source

Journal of Healthcare Engineering

Issue

Vol. 2018, Issue 2018 (31 Dec. 2018), pp.1-12, 12 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2018-10-28

Country of Publication

Egypt

No. of Pages

12

Main Subjects

Public Health
Medicine

Abstract EN

Out-of-hospital cardiac arrest (OHCA) is recognized as a global mortality challenge, and digital strategies could contribute to increase the chance of survival.

In this paper, we investigate if cardiopulmonary resuscitation (CPR) quality measurement using smartphone video analysis in real-time is feasible for a range of conditions.

With the use of a web-connected smartphone application which utilizes the smartphone camera, we detect inactivity and chest compressions and measure chest compression rate with real-time feedback to both the caller who performs chest compressions and over the web to the dispatcher who coaches the caller on chest compressions.

The application estimates compression rate with 0.5 s update interval, time to first stable compression rate (TFSCR), active compression time (TC), hands-off time (TWC), average compression rate (ACR), and total number of compressions (NC).

Four experiments were performed to test the accuracy of the calculated chest compression rate under different conditions, and a fifth experiment was done to test the accuracy of the CPR summary parameters TFSCR, TC, TWC, ACR, and NC.

Average compression rate detection error was 2.7 compressions per minute (±5.0 cpm), the calculated chest compression rate was within ±10 cpm in 98% (±5.5) of the time, and the average error of the summary CPR parameters was 4.5% (±3.6).

The results show that real-time chest compression quality measurement by smartphone camera in simulated cardiac arrest is feasible under the conditions tested.

American Psychological Association (APA)

Meinich-Bache, Øyvind& Engan, Kjersti& Birkenes, Tonje Søraas& Myklebust, Helge. 2018. Real-Time Chest Compression Quality Measurements by Smartphone Camera. Journal of Healthcare Engineering،Vol. 2018, no. 2018, pp.1-12.
https://search.emarefa.net/detail/BIM-1187480

Modern Language Association (MLA)

Meinich-Bache, Øyvind…[et al.]. Real-Time Chest Compression Quality Measurements by Smartphone Camera. Journal of Healthcare Engineering No. 2018 (2018), pp.1-12.
https://search.emarefa.net/detail/BIM-1187480

American Medical Association (AMA)

Meinich-Bache, Øyvind& Engan, Kjersti& Birkenes, Tonje Søraas& Myklebust, Helge. Real-Time Chest Compression Quality Measurements by Smartphone Camera. Journal of Healthcare Engineering. 2018. Vol. 2018, no. 2018, pp.1-12.
https://search.emarefa.net/detail/BIM-1187480

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-1187480