Data Analysis and Accuracy Evaluation of a Continuous Glucose-Monitoring Device

Joint Authors

Cai, Lijun
Ge, Wancheng
Zhu, Zhigang
Zhao, Xueling
Li, Zhanhong

Source

Journal of Sensors

Issue

Vol. 2019, Issue 2019 (31 Dec. 2019), pp.1-8, 8 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2019-12-31

Country of Publication

Egypt

No. of Pages

8

Main Subjects

Civil Engineering

Abstract EN

This study was aimed at analyzing data and evaluating the accuracy of a new subcutaneous continuous glucose-monitoring device by referring to finger-pricking measurement.

The data were obtained from 7 diabetic patients.

An improved implanted flex sensor was used to measure the interstitial glucose concentration every 3 min within 6 days, and five finger-pricking samples were collected every day for comparison.

A periodic glucose change happened every day.

2.45% of CGM values were in the hypoglycemic range (<70 mg/dl), 55.4% were in the normal range (70-180 mg/dl), and 42.15% were in the hyperglycemic range (>180 mg/dl).

The interstitial glucose concentrations (n=204) were well linearly correlated with the capillary glucose concentrations (r=0.94, P<0.01), but a delay occurred between 10 and 40 minutes within two measurements (CGM later), and the individual average delay times were relatively close to 24.6 minutes.

Clarke’s error grid analysis showed that 86.7% of the points fell in zone A, 12.7% fell in zone B, and only 0.6% fell in zone D.

An overall MARD was 10.22%.

This study demonstrated that the CGM was accurate and highly reliable; thus, it constituted a new device for continuous glucose monitoring in diabetic patients.

American Psychological Association (APA)

Cai, Lijun& Ge, Wancheng& Zhu, Zhigang& Zhao, Xueling& Li, Zhanhong. 2019. Data Analysis and Accuracy Evaluation of a Continuous Glucose-Monitoring Device. Journal of Sensors،Vol. 2019, no. 2019, pp.1-8.
https://search.emarefa.net/detail/BIM-1191232

Modern Language Association (MLA)

Cai, Lijun…[et al.]. Data Analysis and Accuracy Evaluation of a Continuous Glucose-Monitoring Device. Journal of Sensors No. 2019 (2019), pp.1-8.
https://search.emarefa.net/detail/BIM-1191232

American Medical Association (AMA)

Cai, Lijun& Ge, Wancheng& Zhu, Zhigang& Zhao, Xueling& Li, Zhanhong. Data Analysis and Accuracy Evaluation of a Continuous Glucose-Monitoring Device. Journal of Sensors. 2019. Vol. 2019, no. 2019, pp.1-8.
https://search.emarefa.net/detail/BIM-1191232

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-1191232