Are tubes containing sodium fluoride still needed for the measurement of blood glucose in hospital laboratory practice ?

Joint Authors

al-Kindi, Manal
Mula Abid, Wad Allah
al-Kharusi, Amirah
al-Lawati, Nuha

Source

Oman Medical Journal

Issue

Vol. 29, Issue 6 (31 Dec. 2014), pp.404-407, 4 p.

Publisher

Oman Medical Specialty Board

Publication Date

2014-12-31

Country of Publication

Oman

No. of Pages

4

Main Subjects

Medicine

Abstract EN

Objectives: To compare glucose values obtained using sodium flouride (NaF) tubes and serum separation tubes (SST) in a tertiary care hospital laboratory setting.

Methods: This study was conducted at the Clinical Biochemistry Laboratory, Royal Hospital, Oman.

During the study period (1 September–30 November 2013), 50 pairs (one NaF tube and one SST) of patient’s blood specimens were randomly collected.

Following separation of plasma (NaF tubes) and serum (SST), glucose concentrations were measured by hexokinase assay using the Architect c8000.

Fifteen pairs of these tubes were kept in the refrigerator at 4°C and plasma/serum glucose concentrations were measured daily up to seven days after collection.

Results: Comparing plasma (NaF) and serum (SST) results of glucose values (n=50) showed an average difference of 0.00mmol/L (range -0.60 to +0.60mmol/L).

Bland Altman analysis gave a non-significant constant bias of 0.10 ±0.195mmol/L (bias ±SD).

Pearson correlation between plasma (NaF) and serum (SST) glucose concentrations revealed a significant correlation approaching unity with r2= 0.9991.

No significant differences in glucose values were noted for both plasma and serum in 15 pairs of NaF and SST tubes when analyzed seven days following refrigeration.

Hemolysis was observed in five (10%) NaF tubes compared with two (4%) SST.

Conclusion: There is no difference in glucose values collected from plasma NaF tubes or serum SST, and so SST can be used in hospital laboratory settings as there are practical advantages, including cost-effectiveness and reduction in blood volume drawn when utilizing these tubes for glucose and other tests from a single blood collection tube

American Psychological Association (APA)

al-Kharusi, Amirah& al-Lawati, Nuha& al-Kindi, Manal& Mula Abid, Wad Allah. 2014. Are tubes containing sodium fluoride still needed for the measurement of blood glucose in hospital laboratory practice ?. Oman Medical Journal،Vol. 29, no. 6, pp.404-407.
https://search.emarefa.net/detail/BIM-831860

Modern Language Association (MLA)

al-Kharusi, Amirah…[et al.]. Are tubes containing sodium fluoride still needed for the measurement of blood glucose in hospital laboratory practice ?. Oman Medical Journal Vol. 29, no. 6 (2014), pp.404-407.
https://search.emarefa.net/detail/BIM-831860

American Medical Association (AMA)

al-Kharusi, Amirah& al-Lawati, Nuha& al-Kindi, Manal& Mula Abid, Wad Allah. Are tubes containing sodium fluoride still needed for the measurement of blood glucose in hospital laboratory practice ?. Oman Medical Journal. 2014. Vol. 29, no. 6, pp.404-407.
https://search.emarefa.net/detail/BIM-831860

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references : p. 407

Record ID

BIM-831860